Chapter 2
INSTALLATION PROCEDURE
Thermo Fisher Scientific
MR23i
2-5
It is not necessary to orientate the rotor relative to the drive shaft in
order to achieve locking. The AUTOLOCK rotor mounting system
allows rotors to be placed in any orientation.
YOU ARE STRONGLY ADVISED TO USE SEALED LIDS ON
ROTORS IN ORDER TO BE PROTECTED AGAINST THE BIOLO-
GICAL HAZARDS RESULTING FROM BROKEN OR LEAKING
TUBES AND BOTTLES.
ROTOR REMOVAL
For rotors with accessible centres press a thumb against the spring
locking device and lift with the fingers while pulling with the other
hand supporting the rotor.
For rotors with inaccessible centres, press the unlocking device
(stored beneath the centrifuge lid) into the centre of the rotor and
then lift out the rotor using both hands. Replace the device beneath
the lid.
SEALED ROTORS MAY BE REMOVED WITH THE LIDS STILL IN
POSITION AND MOVED TO A SAFETY CABINET FOR MANIPU-
LATION OF HAZARDOUS MATERIAL.
Содержание 11174729
Страница 1: ...P N 50116911 1 Analyze Detect Measure Control Instruction Manual MR 23i ...
Страница 2: ......
Страница 3: ...PN 50116911 1 Issued 06 08 Instruction Manual Jouan MR 23i ...
Страница 4: ......
Страница 14: ...Chapter 2 INSTALLATION PROCEDURE 2 6 MR23i Thermo Fisher Scientific ...
Страница 26: ...Chapter 4 OPERATING PRINCIPALES 4 6 MR23i Thermo Fisher Scientific ...
Страница 54: ...Chapter 6 HAZARDS PRECAUTIONS AND LIMITATIONS OF USE 6 18 MR23i Thermo Fisher Scientific ...
Страница 58: ...Annex A GUARANTEE TERMS A 2 MR23i Thermo Fisher Scientific ...
Страница 62: ...Annex B Declaration of conformity B 4 Heraeus Biofuge Primo R Thermo Fisher Scientific ...
Страница 64: ...Index I 2 MR23i Thermo Fisher Scientific ...
Страница 65: ......