Model 3980 and 3986 Incubators _____________________________________________Service
5-6
The infinite heater control is located in the right side of
the incubator top compartment behind the control panel door.
The heater varies the amount of door heat from no heat (zero)
to full heat (100) as indicated by the control dial face. If the
knob is turned past zero, a “click” will indicate that all power
to the door is shut off. If turned past 100, a similar “click” will
indicate that the heat is set at the maximum.
Initially the units leave the factory with the dial set at
40. If desired, the amount of heat can be reduced until
moisture appears on the door, then the heat advanced.
However, in fluctuating ambient conditions, it is
recommended that a minimum of 40% door heat be used.
5.7 Replacing the Air Pump(s)
De-energize all potential sources of energy to
this unit and lockout/tagout their controls.
(O.S.H.A. Regulation, Section 1910-147.)
1. Access to the air pump(s) is obtained by opening the front of the control panel.
Grasp the left side of the control panel housing and pull it gently but firmly, up and
out.
2. The air pump(s) are located midway back toward the right side of the component
area.
3. Remove the screws securing the air pump bracket(s). Remove bracket(s) and set
aside.
4.
Remove the clear outlet tubing from defective air pump(s).
5.
Electrically disconnect air pump(s) from main wiring harness.
5.8 Cleaning and Adjusting the Humidity Steam Generator, P/N 505087
Materials Required:
6-foot stepladder
Flat and Phillips screwdrivers
Clamp-on style Ammeter
Laboratory disinfectant
Siphon, sponge, and cleaning materials
Содержание 3980
Страница 2: ......
Страница 47: ...Model 3980 and 3986 Incubators ______________________________________Specifications 6 3 ...
Страница 49: ......
Страница 50: ......
Страница 51: ......
Страница 52: ......
Страница 53: ......
Страница 54: ......
Страница 55: ......
Страница 56: ......
Страница 57: ......
Страница 58: ......