System Shutdown, Startup, and Reset
__________________________________________ Placing the System in Standby Condition
__________________ LCQ Deca XP Plus Hardware Manual __________________
3-3
3.2 Placing the System in Standby
Condition
The XP Plus system does not need to be shut down completely if you are not
going to use it for a short period of time, such as overnight or over weekends.
When you are not going to operate the system for 12 h or more, you can leave
the system in a Standby condition.
Use the following procedure to place the XP Plus system in the Standby
condition:
1. Wait until data acquisition, if any, is complete.
2. Turn off the flow of sample solution from the LC (or other sample
introduction device).
3. From the Tune Plus window, choose
Control | Standby
(or click on the
On/Standby button) to put the MS detector in Standby condition
.
When
you choose Control | Standby, the XP Plus turns off the electron
multiplier, conversion dynode, 8 kV power to the API source, ring
electrode RF voltage, and octapole RF voltage. The XP Plus also turns off
the auxiliary gas and sets the sheath gas flow to 0 units. See Table 3-1 on
page 3-10 for the On/Off status of MS detector components when the MS
detector is in the Standby condition. The System LED on the front panel
of the MS detector is illuminated yellow when the system is in Standby
condition.
4. Flush the spray shield and the entrance end of the ion transfer capillary of
the API source as describe in the topic Flushing the Spray Shield, Ion
Sweep Cone, Spray Cone, and Ion Transfer Capillary in the User
Maintenance chapter. Cap the ion transfer capillary with the septum.
Leave the API flange withdrawn from the spray shield.
5. Purge the rotary-vane pump oil as described in the topic Purging the Oil
in the Rotary-Vane Pump in the System Shutdown, Startup, and
Reset chapter.
6. Leave the MS detector power On.
7. Leave the LC power On.
8. Leave the autosampler power On.
9. Leave the data system power On.
Note. For instructions on how to operate the LC from the front panel, refer
to the manual that came with the LC.