___________________________________________ Replacing the RF Voltage Control PCB
__________________ LCQ Deca XP Plus Hardware Manual _________________
5-61
10. Reconnect the two coaxial cables to the BNC connectors that are located
on the rear of the Low Pass Filter PCB.
11. Reinstall the rear cover of the MS detector as follows:
a.
Place the cover against the rear of the MS detector such that the
screws in the MS detector chassis insert into the guide slots on the
rear cover.
b. Slide the rear cover down about 1.2 cm (0.5 in.) until the screws lock
in the guide slots on the cover.
c.
With a Phillips screwdriver, tighten the ten screws that secure the rear
cover to the chassis of the MS detector.
12. Reinstall the top cover of the MS detector as described in the topic
Reinstalling the Top Cover of the MS Detector on page 5-16.
13. Place the electronics service switch in the Operating Position.
14. Run the Tune Plus diagnostics to verify that the system is operational.