Service Maintenance
______________________________________ Replacing PCBs and Assemblies in the Tower
__________________ LCQ Deca XP Plus Hardware Manual _________________
5-45
7. Unseat the PCB from the backplane and pull it out of the embedded
computer.
8. Unpack the new PCB. Retain the packing materials so that you can pack
and ship the defective PCB to the Thermo Finnigan Repair Center in San
Jose. Be sure to note the apparent problem or symptoms on the
enclosed forms.
9. Seat the new PCB in the backplane.
10. With a Phillips screwdriver, reinstall the screw that holds the PCB to the
card cage.
11. Reconnect all electrical cables to the PCB that you replaced.
12. Reinstall the front cover of the embedded computer. Tighten by hand the
two fasteners that hold the front cover to the embedded computer.
13. Close the front doors to the MS detector.
14. Place the electronics service switch in the Operating Position.
15. Run the Tune Plus diagnostics to verify that the system is operational.
Replacing the Vent Delay PCB and Backup
Battery, Ion Gauge, and Vent Valve
The Vent Delay PCB, ion gauge, and vent valve can be accessed from the top
of the MS detector. See Figure 5-15.
Caution. To prevent damage to the electronics due to electrostatic
discharge, attach an electrostatic discharge (ESD) strap to your wrist before
continuing.
CAUTION. The XP Plus system must be shut down and the power cord
unplugged before you service the vent valve, Vent Delay PCB, Convectron
gauge, or ion gauge.