Installation & Wiring
2-6
Polysonics SX40 User Guide
Thermo Electron Corporation
Figure 2–7.
AC power connections
Later versions of the instrument (those with P/N 22686-0002 Rev. G and
later) have DC operation built into the instrument. Refer to the following
instructions and Figure 2–8.
1.
Remove the meter access cover.
2.
Locate the power input block, and turn the terminal screws on top of
the input block counterclockwise to fully open the wire connections.
3.
Make the following connections:
a.
+24VDC to P2 +24
b.
24 VDC return to either of the P2 GND connections
4.
Turn the terminal screws clockwise to close them connections.
DC Power