2 - 2
2.3
Symptom: Unit will not communicate with alarm system (Remote Alarm Contacts/Analog
System/Serial communication)
Order Possible
Cause
Verification
Action
Check that all connections are tight and properly
wired. Missing wires?
Tighten and / or correct wiring.
1
Remote Alarm Contacts
Measure continuity of alarm contacts between
pins 6 & 7 and 6 & 8 while toggling in test mode
(pin 6 is common).
Replace micro board if contacts do not change
state.
Verify the setting is correct in hidden calibration
and corresponds to the dip switch settings on the
micro board per the electrical schematic.
Make changes so the dip switch settings and hidden
calibration settings match.
Measure output. Compare with
Analog Output
Reference Table
(page 3-1).
Call the factory service department 888-213-1790.
See section 4.2.p Analog Calibration (page 4-3).
2
Analog Output
Check that all connections are tight and properly
wired. Missing wires?
Tighten and / or correct wiring.
Verify address is correct in configuration menu
and matches the 1535 address.
Change so that the unit address of the freezer
matches the 1535 address.
Verify the setting is correct in hidden calibration
and corresponds to the dip switch settings on the
micro board per the electrical schematic.
Make changes so that the dip switch settings and
hidden calibration settings match.
Check that all connections are tight and properly
wired. Missing wires?
Tighten and / or correct wiring.
Tighten and / or correct wiring.
3
Serial Communication
Verify that the cable is a DB9 (straight through)
extension cable.
Replace cable if necessary.