Environmental Chamber ___________________________________________________________________________Service

5 - 5

WATLOW 982 CONFIGURATION RECORD

CUSTOMER:

3920 & 3940 (RELEASE 9)

JOB NUMBER:
SERIAL NUMBER:
CONTROL TYPE: TEMPERATURE (RTD)
PREPARED BY:

GLS

DATE:

COMPLETED BY:

DATE:

Switch Configuration:

(1)*** Main Boards

Input 1 Board

SW1

ON

SW1

OFF

SW2

ON

SW2

OFF

SW3

ON

Software Configuration:

Setup Menus

(2)**** gLbL: C_F

C

Anun

OFF

PtyP

ti

PStr

StPt

Err

nLA

LoP

-100

gSd

0.0

LOC

2*

Ei1

NO

HiP

100

Pout

Cont

Ei2

NA

AtSP

90

IdSP

NA

*LOC should be set at 0 until factory testing and calibration is complete

(3)**** InPt:

In1

RT.D***

CAL1

**

rL2

NA

CAL2

NA

dEC1

NA

rtd1

DIN

rH2

NA

Hunt

NA

rL1

-20.0

Ftr1

2

LrnL

NA

rH1

80.0

In2

NA

LrnH

NA

**This is a calibration factor and will vary from unit to unit.

(4)**** OtPt: Ot1

Ht

AL2

NA

LAt3

NA

SIL4

OFF

PrC1

NA

LAt2

NA

SIL3

NA

Aout

Prc1

HyS1

0.1

SIL2

NA

Ot4

AL4n

PrC3

4-20

Ot2

CL

Ot3

NA

AL4

Pr1

ArL

-20.0

PrC2

NA

AL3

NA

HyS4

0.1

ArH

80.0

HyS2

0.1

HyS3

NA

LAt4

nLA

ACAL

**

**This is a calibration factor and will vary from unit to unit.

(5)**** COM: bAUd

9600

dAtA

7o

PrOt

FULL

Addr

0

IntF

NA

Operation Menus

(6)**** SYS: Ei1S

NA

Ent4

NA

A3LO

NA

A4Hi

34.0

Ei2S

NA

A2LO

NA

A3LHi

NA

Aut

OFF

Ent3

NA

A2Hi

NA

A4LO

-73.3

(7)**** PID:

Pb1

2.0

dE1

NA

It2

NA

dB

0.0

rE1

0.50

Ct1

5.0

rA2

0.25

It1

NA

Pb2

1.0

dE2

NA

rA1

0.19

rE2

0.25

Ct2

30.0

PROG: See Programming Sheets if required.

***SET THESE PARAMETERS FIRST
****CONFIGURE CONTROL IN IDENTIFIED ORDER.

11/05/03

Содержание 3920

Страница 1: ...Model 3920 Forma Environmental Chamber Operating and Maintenance Manual Manual No 7043920 Rev 3 ...

Страница 2: ...it describes are subject to change with out notice Thermo Electron makes no representations or war ranties with respect to this manual In no event shall Thermo be held liable for any damages direct or incidental arising out of or related to the use of this manual MANUAL NUMBER 7043920 3 22678 IN 3229 1 17 05 Removed 270129 from terminals 15 and 16 temp control aks 2 22454 IN 3208 9 14 04 Watlow SD...

Страница 3: ...procedimientos asociadas con este simbulo Equipment being maintained or serviced must be turned off and locked off to prevent possible injury Risques potentiels liés à l énergie L équipement en entretien ou en maintenance doit être éteint et mis sous clé pour éviter des blessures possibles Geräte an denen Wartungs oder Servicearbeiten durchgeführt werden müssen abgeschaltet und abgeschlossen wer d...

Страница 4: ...Environmental Chamber _________________________________________________________________________Service iii ...

Страница 5: ...el can provide precise calibrations as a billable service at your location Calibration after a warranty repair is covered under the warranty Damage resulting from use of improper quality water chemicals or cleaning agents detrimental to equipment materials Service calls for improper installation or operating instructions Corrections to any of the following are billable services 1 electrical servic...

Страница 6: ...00140 IR CO2 Option 2 3 a Connecting the CO2 Source Figure 2 4 2 3 b Setting the CO2 Content 2 4 c CO2 Control and Indicators 2 4 Section 3 Start Up and Operation 3 1 3 1 The Control Panel 3 1 3 2 Setting the Operating Temperature Figure 3 4 3 2 3 3 Air Exchange Ventilator Caps 3 2 3 4 Heatless Dryer P N 1900139 optional 3 2 Section 4 Routine Maintenance 4 1 4 1 Cleaning the Incubator 4 1 Section ...

Страница 7: ......

Страница 8: ...adequately describe such evi dence of loss or damage may result in the carrier refusing to honor a damage claim Immediately call the delivering carrier and request that their representative perform an inspection Do not discard any of the packing material and under no circum stances move the shipment from the receiving area 1 3 Responsibility for Shipping Damage For products shipped F O B Marietta ...

Страница 9: ......

Страница 10: ...nnel by locating the rivet into a slotted hole far end first Pull the channel toward the front and slide the front rivet on the channel into the slotted hole and press down Make sure that the channels are oppo site each other so that the installed shelf will be level d Leveling the Unit Place a bubble type level on a shelf inside the incubator Adjust the feet as needed counterclockwise to lengthen...

Страница 11: ...ted before normal operation under the main temperature controller can be resumed Note When the chamber temperature control setpoint is changed the overtemp safety thermostat must be reset to accom modate the change Note The overtemp control is not directly calibrated The num bers on the dial are for reference only 2 7 Setting the Undertemp Safety Thermostat Allow the chamber temperature and humidi...

Страница 12: ... is capable of recording both temperature and humidity and printing alphanumeric chart data on blank heat sensitive chart Refer to the Honeywell Configuration Record at the end of Section 5 and the supplemental Honeywell Recorder manual 2 10 IR CO2 Option This section applies to units with the IR CO2 option only a Connecting the CO2 Source Figure 2 5 For the most economical use the liquid CO2 supp...

Страница 13: ...Indicators Sample Port Figure 2 6 The sample port is used for checking CO2 percentage in the incubator chamber by an independent test instrument such as with a Fyrite or similar CO2 test instrument To prevent CO2 loss the sample port must be capped when it is not in use CO2 Alarm Figure 2 6 The CO2 alarm is factory set to activate when the chamber CO2 content deviates from configured alarm set poi...

Страница 14: ...r to the refriger ation system The refrigeration indicator lights when the refrigeration switch is on and the compressor is receiving power Defrost Switch and Indicator Light Figure 3 3 The defrost switch controls power to the defrost sys tem Setting the defrost switch to Auto will provide two 15 minute defrost cycles during a twenty four hour period The defrost indicator lights when the defrost s...

Страница 15: ...ers and maintain the incubator s cabinet temperature at the overtemp safety control point Note The overtemp control is not directly calibrated The num bers on the dial are for reference only If an overtemp condition occurs the alarm can only be silenced by raising the overtemp safety thermostat setting However the cause of the problem must be determined and cor rected before normal operation under...

Страница 16: ...r with soap and water or with any non abrasive commercial spray cleaner Clean the inside of the chamber with alcohol and or soap and water Disinfect the interior panels with a general use laboratory disinfectant diluted according to the manufacturer s instruc tions Rinse the surface thoroughly after each cleaning and wipe the surfaces dry Always rub in the direction of the finish pol ish lines Do ...

Страница 17: ......

Страница 18: ...st of preventive maintenance requirements See the specified section of the instruction manual for further details Thermo Electron has qualified service technicians using NIST traceable instruments available in many areas For more information on Preventive Maintenance or Extended Warranties please contact us at the number below Cleaning and calibration adjustment intervals are dependent upon use en...

Страница 19: ...n Experiments side panels with disinfectant Rinse everything well with sterile distilled water More frequent decontamination may be required depending on use and environmental conditions 5 4 5 7 Verify and document all calibrations at the minimum Clean drip pan and drain lines Clean refrigeration system condenser Verify defrost cycle for below 10 C operation Change filters under normal conditions ...

Страница 20: ...g probe 5 3 Replacing the Temperature Sensor 1 Locate the probe mounting plate in the center of the right side of the incubator interior 2 Open the mounting plate by removing the screws that hold it in place 3 Locate the humidity sensor mounted on the inside of the panel in a black housing Note the angle of the probe 4 Grasp and unplug the probe from the probe cable 5 When replacing the humidity s...

Страница 21: ... displays show 2 Loc Figure 5 5 Press the Down arrow key until the 2 becomes 0 zero as shown in Figure 5 6 c Controller Configuration The Watlow 982 Temperature Controller has been config ured at the factory Copies of the Watlow Configuration records are included at the end of this section Do not reconfigure the controller without first consulting the Technical Services department at 888 213 1790 ...

Страница 22: ...ntal Chamber _____________________________________________________________________________Service 4 Turn off the main power switch 5 Wear a grounding wrist strap to avoid damaging any of the electrical components 6 Remove the 982 controller module s by pressing in the four retaining tabs two on the right side two on the left side Pull the controller module out by gently rocking it from side to sid...

Страница 23: ...o Figure 5 9 Then adjust as necessary If an adjustment is necessary press either the 1 or 2 button to move the pen left or right The button must be held about five seconds before the pen begins to move Release the button when the pen matches the thermometer 5 6 Setting the Door Heater Control High voltage is present behind control panel Servicing must be performed only by qualified electrical serv...

Страница 24: ...A CAL2 NA dEC1 NA rtd1 DIN rH2 NA Hunt NA rL1 20 0 Ftr1 2 LrnL NA rH1 80 0 In2 NA LrnH NA This is a calibration factor and will vary from unit to unit 4 OtPt Ot1 Ht AL2 NA LAt3 NA SIL4 OFF PrC1 NA LAt2 NA SIL3 NA Aout Prc1 HyS1 0 1 SIL2 NA Ot4 AL4n PrC3 4 20 Ot2 CL Ot3 NA AL4 Pr1 ArL 20 0 PrC2 NA AL3 NA HyS4 0 1 ArH 80 0 HyS2 0 1 HyS3 NA LAt4 nLA ACAL This is a calibration factor and will vary fro...

Страница 25: ...0 10 Aut OFF Pbht 1 0 A2hı 20 04 dSPL Display Intensity 5 to 99 rESt Restore Factory Calibration Select YES to restore Usrr Restore User Settings Select YES to restore user configuration USrS Save User Settings Select YES to save user configuration dFLt Default Parameters Select YES to reset to default configuration NOTE Not all available parameters are listed on this page See Watlow manual for mo...

Страница 26: ...ss steel Exterior Cold rolled steel Insulation 2 5 1cm Foamed urethane Outer Door Gasket Four sided vinyl compression Finish Powder coated Salt spray tests exceed 1000 hrs per ASTM Standard B117 85 Weights Net 700 lbs Shipping Motor 850 lbs Temperature Alarm Sensor Thermostat Controller Thermostat Setpoint Analog reference dial Alarm Audible visual Fittings Drain Port 3 8 OD Copper Unit Heat Load ...

Страница 27: ......

Страница 28: ...y 1 290162 RH Sensor 1 231221 Watlow 982 Controller 1 231224 Watlow SD Controller 1 270129 Arc filters for Watlow 982 controller 1 320277 Triac 25A 2 285614 20A Fuse 2 137022 Heater steam 1 430940 Over Temp Thermostat 1 430941 Under Temp Thermostat 1 410048 Defrost Timer 1 630090 Heater wirewound 1 205062 Condensing Unit 1 4 HP 115V R 134A 1 991334 Solenoid 120V 101 Port 1 230214 Thermal Fuse 250V...

Страница 29: ......

Страница 30: ...Environmental Chamber ______________________________________________________________Refrigeration Schematic 8 1 ...

Страница 31: ......

Страница 32: ...Environmental Chamber __________________________________________________________________Electrical Schematics 9 1 ...

Страница 33: ...Environmental Chamber __________________________________________________________________Electrical Schematics 9 2 ...

Страница 34: ...Environmental Chamber __________________________________________________________________Electrical Schematics 9 3 ...

Страница 35: ......

Страница 36: ... warranty period The Technical Services Department must give prior approval for return of any components or equipment At Thermo s option all non conforming parts must be returned to Thermo Electron Corporation postage paid and replacement parts are shipped FOB destination THIS WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES WHETHER WRITTEN ORAL OR IMPLIED NO WARRANTIES OF MERCHANTABILITY...

Страница 37: ...der this warranty shall not extend the warranty to either the equipment or to the component part beyond the original warranty period The Technical Services Department must give prior approval for return of any components or equipment At Thermo s option all non conforming parts must be returned to Thermo postage paid and replacement parts are shipped FOB destination THIS WARRANTY IS EXCLUSIVE AND I...

Страница 38: ...Thermo Electron Corporation Controlled Environment Equipment Millcreek Road P O Box 649 Marietta Ohio 45750 U S A Telephone 740 373 4763 Telefax 740 373 4189 ...

Отзывы:

Похожие инструкции для 3920