UM-35-SG manual (UM09)

110/120 VAC 

Selection header

200/240 VAC 

Selection header

Component Layout

Calibration Procedure

Decimal Point Selection

1.XXX
1X.XX
1XX.X
1XXX

Cut loop to turn 

OFF last digit

0.56" Display

Decimal Select 

Header

Optional Isolated auto-sensing 

AC/DC 9 to 36V DC / 12 to 24V AC

P.N.: PS7

Connector Pinouts

This meter uses plug-in type screw terminal connectors for all 

input and output connections� The power supply connections 

(pins 14 and 15) have a unique plug and socket outline to 

prevent cross connection� The main board uses standard right-

angled connectors�

 

 

14 15

10 11

1 2 3 4

12

AC/DC Power

ZERO

SP

AN

HOLD

INPUT Hi

INPUT Lo

+ EXC

EXC

TEST

COMMON

AC 

Neutral

– DC

AC 

Line

+ DC

Signal Input Pin

 

This meter uses plug-in type 

screw terminal connectors for 

all connections�

Connectors

!

WARNING

AC and DC input signals and power supply voltages can be 

hazardous. Do Not connect live wires to screw terminal plugs, and do not 

insert, remove or handle screw terminal plugs with live wires connected.

Signal Conditioning Components

SPAN Potentiometer (Pot)

The 15 turn SPAN pot is always on the right 

side (as viewed from the front of the meter). 

Typical adjustment is 100% of the input sig-

nal range.

ZERO Potentiometer (Pot)

The Optional ZERO pot when installed is 

always to the left of the SPAN pot (as viewed 

from the front of the meter). Typically it 

enables the displayed reading to be offset 

±100 counts.

Optional Face Plate Descriptors

Decimal selection 

is made by moving 

the jumper to the 

indicated position 

on the header 

for the decimal 

required on the 

front of the display 

board�

To open meter,

 insert 

a flat head screwdriver 

or similar instrument in 

both slots on the side 

of the cover and pry 

open� The UM-Series 

meters slide out from 

the front of the case as 

a complete assembly�

To customize the face plate, clear 

adhesive label containing various 

popular descriptors may be ordered� 

Choose the descriptor desired, peel 

off the adhesive backing and align 

the descriptor in the center right of 

the faceplate� 

P�N�: DU-CASEDES

1�  Select the appropriate position on the INPUT RANGE 

SELECT Header, to correspond with the mV/V output of the 

transducer being used� 

2�  Connect the transducer to the UM-35-SG as shown in the 

con nec tion  diagram�

3�  With no pressure or load being applied to the transducer, 

adjust the ZERO pot so that the display reads 000�

4�  Apply a known input, and adjust the SPAN pot until the 

meter displays the required reading for the pressure or load 

being applied�

�  The UM-35-SG meter is now calibrated and ready for use� 

SPAN Pot

ZERO Pot

mV/V Select Header

(2mV/V or 20mV/V)

Span Pot Hi/Low

Adjustment Range 

Header

Excitation Range 

Header (5V or 10V)

Span Pot Adjustment Range Header

 

Low : 0-1000 Counts
Hi : 1001-1999 Counts

Range

HI

LO

HI

LO