28
2. OPERATION
2-1. Basic Operation
The LSG Series supports 7 main operating modes:
CC, CC+CV, CR, CR+CV ,CV ,CP, CP+CV
2-1-1. CC Mode
Description
In Constant Current Mode the load units will sink
the amount of current programmed. Regardless
of the voltage, the current will stay the same. For
more details on CC mode, please see the
appendix on page 133.
Warning
If you change the mode or the range when the
load is already on, the load will be turned off
automatically.
Operation
1.
Make sure the load is off.
2.
Press
Main
.
3.
Select CC mode with the
Mode[F1]
soft-key.
4.
Select the current range with the
I Range[F2]
soft-key.
Range:
High, Middle, Low
5.
Select the voltage range with the
V Range[F3]
soft-key.
Range:
High, Low
6.
Set the current level parameters using the scroll
wheel and number pad.
For Static mode, set
CC A Value
and/or
CC B
Value
.
For Dynamic mode, set
Level1
and
Level2
.
The maximum and minimum current levels
depend on the selected ranges.
7.
To add CV mode to CC mode (CC+CV), see
page 33.
8.
Set the remaining basic configuration settings
such as the slew rate, and switching mode
settings. See page 37 for details.