10
2-6. Select Voltage range
Auto range
To turn the automatic range selection On/Off, press
the AUTO key.
Manual range
Press the Up or the Down key to select the range.
AUTO indicator turns Off automatically. If the
appropriate range is unknown, select the highest
range.
Selection list
Range
Resolution / Full scale @ slow rate
Resolution
Full scale
100mV
1µV
120.000mV
1V
10µV
1.20000V
10V
100µV
12.0000V
100V
1mV
120.000V
750V (AC)
10mV
750.00V
1000V(DC, AC+DC)
10mV
1000.0V
Note
For more detailed parameters, see the specifications at page49.
2-7. Voltage conversion table
This table shows the relationship between AC, DC, and AC+DC reading in various waveforms.
Waveform
Peak to Peak
AC
(True RMS)
DC
AC + DC
(True RMS)
Sine
PK-PK
2.828
1.000
0.000
1.000
Rectified Sine (full
wave)
PK-PK
1.414
0.435
0.900
1.000
Rectified Sine (half
wave)
PK-PK
2.000
0.771
0.636
1.000
Square
PK-PK
2.000
1.000
0.000
1.000
Rectified Square
PK-PK
1.414
0.707
0.707
1.000
Rectangular Pulse
PK-PK
X
Y
2.000
2K
K=
)
2
(
D
D
D=X/Y
2D
D=X/Y
2
D
D=X/Y
Triangle Sawtooth
PK-PK
3.464
1.000
0.000
1.000
Содержание DL-2052
Страница 1: ...B71 0422 01 INSTRUCTION MANUAL DIGITALMULTIMETER DL 2052 ...
Страница 55: ...48 8 3 Status system ...
Страница 60: ...53 9 13 External Dimensions Figure ...