Input Signal
Vin
DMM
(DC_Amp)
OUT
IN
SET
10
I
V
R
=
Basic EVM Configuration
www.ti.com
3.2
Current Mode Configurations
3.2.1
Single-Ended Output
The first current-mode configuration creates a single-ended output from a single-ended input. The jumper
settings for this configuration are listed in
Table 6
. Connect your load to TB4 between OUT+ and OUT–.
Apply ±15 V to V+ and V–. Connect a 5-V signal to V
IN
. Measure the current through the load by attaching
your digital multimeter (DMM) between OUT– and GND. The current though the load is approximately 20
mA. The current output is calculated by the transfer function in
Equation 3
:
(3)
The default EVM value for R
SET
is 2.5 k
Ω
.
Table 6. Single Ended Current Output Jumper Settings
Jumper
Function
J1
On
J2
On
J3
On
J4
Off
J5
Off
J6
Off
J7
Off
Figure 6. XTR300 Single-Ended Current Output Configuration Circuit Drawing
8
XTR300EVM
SBOU034A – February 2006 – Revised February 2013
Submit Documentation Feedback
Copyright © 2006–2013, Texas Instruments Incorporated