The feedback resistor for hysteresis on channel A will be the combination of R11 and R13. Both resistors must
be populated. When measuring the INA+ input signal on an oscilloscope, we recommend using a dual channel
function generator so that one of its outputs can be directly connected to the oscilloscope. Ensure that both
signal generator channels are outputting the same signal.
+
–
Input Resistor
OUTA
INA-
INA+
Input Signal
(INA+SENSE)
TLV3602A
Option for
Feedback for
Hysteresis
Option for
Feedback for
Hysteresis
R11
R13
R7
R
FEEDBACK
= R
11
+ R
13
Figure 5-5. TLV3602 Hysteresis Schematic for Channel A
For non-inverting hysteresis on Channel B, the INB+SENSE SMA connection will serve as the input to the
device. The input resistor will be series resistor R8. The feedback resistor for hysteresis on channel B will be
the combination of R9 and R10. Both resistors must be populated. When measuring the INB+ input signal on an
oscilloscope, we recommend following the same recommendations as Channel A.
+
–
Input Resistor
OUTB
INB-
INB+
Input Signal
(INB+SENSE)
TLV3602B
Option for
Feedback for
Hysteresis
Option for
Feedback for
Hysteresis
R9
R10
R8
R
FEEDBACK
= R
9
+ R
10
Figure 5-6. TLV3602 Hysteresis Schematic for Channel B
Board Setup
10
TLV3602EVM User's Guide
SNOU193 – MAY 2022
Copyright © 2022 Texas Instruments Incorporated