Meas_x
Y Axi
s Titl
e (
Unit)
0
2000
4000
6000
8000
10000
12000
14000
16000
18000
20000
-160
-140
-120
-100
-80
-60
-40
-20
0
-142.86dBc (Second Harmonic)
Test Setup and Results
11
SBOU212 – September 2018
Copyright © 2018, Texas Instruments Incorporated
INA1651EVM SoundPlus™ High Common-Mode Rejection Line Receiver
Evaluation Module
4.8
Fast-Fourier Transform (FFT)
The
FFT
measurements were taken with a 1-kHz, 10-V
RMS
signal on the output of the INA1651. A
10 V
RMS
fundamental corresponds to 0 dBc in the
FFT
measurement.
shows an
FFT
of the
INA1651 output. The second harmonic was measured to be –142.86 dBc.
Figure 10. FFT Measurement
1 kHz, 10-V
RMS
Output