![Texas Instruments PGA309 Скачать руководство пользователя страница 61](http://html.mh-extra.com/html/texas-instruments/pga309/pga309_quick-start-manual_1097110061.webp)
SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
61
This is the hardware
configuration that this
ratiometric calibration example
details. In this example, the
Sensor-Emulator-EVM is used
to emulate the bridge sensor
and the Diode. Note that the
device power supply is used to
provide excitation for the
sensor. So for this
configuration, the Vexc pin on
the PGA309 is not used and
consequently, the PGA309
cannot correct for nonlinearity
verses applied stimulus.
Temperature nonlinearities of
span and offset will still be
corrected.
Vexc
Vin+
Vin-
Temp
PGA309
Vsa
5V
GND
Ref_In/
Ref_Out
Real World Inputs
PGA309 Ratiometric
Example