16. To test the results of the calibration click the
Microcontroller State
field to begin running the
microcontroller, and navigate to the
Voltage Monitor
tab in the tool panel.
17. Select the second pressure point in the
Resistive Bridge Configuration
panel, by setting the resistance
of the leg to "5.0806k". The VOUT voltage in the tool panel should match the voltage selected for the
Pressure 2 output during the calibration procedure.
4.13 PGA302 EEPROM Settings Page (PGA302 Only)
illustrates the PGA302 EEPROM settings page.
Figure 19. PGA302 EEPROM Settings Page