Quick Start Setup and Use
This connection method allows the user to either simulate the voltage developed across a sense resistor
based on a given set of system conditions, or to connect the INA210-214EVM remotely to an existing
shunt already included in an example application.
illustrates a measurement configuration without
a shunt.
To configure a measurement evaluation without a shunt, follow these procedures.
1. Connect a differential voltage to the V
IN+
and V
IN–
terminals of T2. With the reference voltage set at
ground, ensure that the V
IN+
terminal is the more positive of the two inputs.
2. Measure the output voltage at the V
OUT
terminal of T2.
Figure 3. Measurement without Shunt
Note:
The output voltage is equal to the gain of the device multiplied by the differential voltage
measured directly at the device input pins.