PF Status Snapshot Data Flash
44
SLUUBD3D – September 2015 – Revised September 2018
Copyright © 2015–2018, Texas Instruments Incorporated
Permanent Fail
Class
Subclass
Name
Type
Min
Max
Default
Unit
Description
PF Status
Device Voltage
Data
Cell Voltage 15
I2
0
32767
0
mV
Cell 15 voltage
PF Status
Device Voltage
Data
Bat Direct Voltage
I2
0
32767
0
mV
Cell stack voltage
4.20.3 Device Current Data
Class
Subclass
Name
Type
Min
Max
Default
Unit
Description
PF Status
Device Current
Data
Current
I2
–32768
32767
0
mA
Current()
4.20.4 Device Temperature Data
Class
Subclass
Name
Type
Min
Max
Default
Unit
Description
PF Status
Device
Temperature
Data
TS1 Temperature
I2
–32768
32767
0
0.1°K
TS1 temperature
PF Status
Device
Temperature
Data
TS2 Temperature
I2
–32768
32767
0
0.1°K
TS2 temperature
PF Status
Device
Temperature
Data
TS3 Temperature
I2
–32768
32767
0
0.1°K
TS3 temperature
4.20.5 AFE Regs
Class
Subclass
Name
Type
Length in Bytes
Min
Max
Default
PF Status
AFE Regs
AFE Sys Stat
H1
1
0x00
0xFF
0x00
PF Status
AFE Regs
AFE Cell Balance 1
H1
1
0x00
0xFF
0x00
PF Status
AFE Regs
AFE Cell Balance 2
H1
1
0x00
0xFF
0x00
PF Status
AFE Regs
AFE Cell Balance 3
H1
1
0x00
0xFF
0x00
PF Status
AFE Regs
AFE Sys Control 1
H1
1
0x00
0xFF
0x00
PF Status
AFE Regs
AFE Sys Control 2
H1
1
0x00
0xFF
0x00
PF Status
AFE Regs
AFE Protection 1
H1
1
0x00
0xFF
0x00
PF Status
AFE Regs
AFE Protection 2
H1
1
0x00
0xFF
0x00
PF Status
AFE Regs
AFE Protection 3
H1
1
0x00
0xFF
0x00
PF Status
AFE Regs
AFE OV Trip
H1
1
0x00
0xFF
0x00
PF Status
AFE Regs
AFE UV Trip
H1
1
0x00
0xFF
0x00
4.21 Black Box Recorder
The Black Box Recorder maintains the last three updates of
SafetyStatus()
in memory. When entering
PERMANENT FAIL mode, this information is written to data flash in addition to the first three updates of
PFStatus()