Battery Management Studio
3.3
Calibration Screen
Calibrate the voltages, temperatures, and currents to provide good gauging performance. Press the
Calibration
button to select the
Advanced Calibration
window. See
Figure 4. Calibration Screen
3.3.1
Voltage Calibration
•
Measure the voltage from BAT+ to BAT– and enter this value in the
Applied Voltage
field and select
the
Calibrate Voltage
box.
•
Press the
Calibrate Gas Gauge
button to calibrate the voltage measurement system.
•
Deselect the Calibrate Voltage boxes after voltage calibration has completed.
3.3.2
Temperature Calibration
•
Enter the room temperature in the
Applied Temperature
field and select the
Calibrate Temperature
box
and select the thermistor to be calibrated. The temperature value must be entered in degrees Celsius.
•
Press the
Calibrate Gas Gauge
button to calibrate the temperature measurement system.
•
Deselect the
Calibrate Temperature
box after temperature calibration has completed.