170 MHz, +15 dBm
Low-Noise
Power Supply
5 VDC @ 3 A
Power
Switch
TSW14J56EVM
PC Running
HSDC Pro
Mini-USB Cable
170-MHz
Bandpass
Filter
USB 3.0 Cable
ADS54J42
EVM
AINP
BINP
Analog Input
Low-Noise
Signal Generator
5-VDC Power Supply Cable
Low-Noise
Power Supply
5 VDC @ 3 A
Quick Start Guide
5
SLAU674 – February 2016
Copyright © 2016, Texas Instruments Incorporated
ADS54J42EVM
2.2
Hardware Setup Procedure
A typical test setup using the ADS54J42EVM and TSW14J56EVM is shown in
. This is the test
setup used for the quick start procedure. The rest of this section describes the hardware setup steps.
Figure 2. Quick Start Test Setup
2.2.1
TSW14J56EVM Setup
First, set up the TSW14J56EVM using the following steps:
1. Connect the ADS54J42EVM to the TSW14J56EVM using the FMC connectors.
2. Connect the included power supply cable to connector J11 (+5V IN) and the other end to a
+5 VDC ±0.3 VDC, 3-A power supply.
3. Connect the included USB cable between the host computer and USB connector (J9).
4. Turn on the power supply. Flip the power switch (SW6) to the
ON
position.