www.ti.com

Basic Test Procedure

2

Basic Test Procedure

This section outlines the basic test procedure for testing the EVM.

2.1

Test Block Diagram

The test set-up for general testing of the ADS42xx/ADS58C28 EVM with the TSW1200 capture card is
shown in

Figure 3

.

Figure 3. Test Set-up Block Diagram

2.2

Verify Board Set-up

Verify jumper settings are in the correct position as outlined in

Table 1

.

Table 1. Default ADS58C28/42xx EVM Rev B Jumper Setting for Serial

Interface

(1)

Jumper

Default position

Function

JP15

Short 1 2

DC supply

JP16

Short 1 2

DC supply

JP17

Short 2 3

DC supply, LDO

JP19

Short 2 3

DC supply, LDO

JP28

Short 2 3

DC supply, LDO

JP29

Short 2 3

DC supply, LDO

JP26

Open

DC supply for ext buffer

JP27

Open

DC supply for ext buffer

JP3

Short 2-3

OPA power down

JP4

Short 2-3

OPA power down

JP22

Open

SDOUT to FPGA

JP20

Short 1 2

CDC

JP21

Short 1 2

CDC

J14

Short 1 2

CDC power down

J18

Open

CDC, VCXO

JP8

Short 2 3

ADC SCLK for SPI

JP9

Short 2 3

ADC SDATA for SPI

JP10

Short 2 3

ADC SEN for SPI

JP11

Short 2 3

ADC for SPI, also reset

JP 12

Short 1 2

ADC Low speed mode disable

JP 13

Open

JP14

Short 7 8

ADC 2’s complement, DDR LVDS

(1)

The EVM schematic shows default setting of JP8 to JP11 as parallel interface (

Table 2

which is for

EVM installation. After EVM tested and released these jumpers are set as serial interface (

Table 1

).

5

SLAU333 – March 2011

ADS42xx EVM

Submit Documentation Feedback

© 2011, Texas Instruments Incorporated

Содержание ADS42 EVM Series

Страница 1: ...ults 8 3 Optional Configurations 8 3 1 THS4509 Input Op Amp Configuration 8 3 2 On Board CDC72010 Clock 9 3 3 Parallel CMOS Output 9 List of Figures 1 ADS58C28_ADS42xx_GUI Front Panel Top Level 3 2 AD...

Страница 2: ...as two register tabs Top Level and Advanced The Top Level tab provides an interface to the most used registers The Advanced tab includes less used registers and provides an option to manually input ad...

Страница 3: ...GUI to bring the advanced panel to the front On this panel user can write address and data of any register into the address data boxes as shown in Figure 2 45 is address and 4 is data Press Enter on t...

Страница 4: ...y updated o Select Load Regs button o Double click on the desired register file o Click on Send All to ensure all of the values are loaded properly 1 2 4 Miscellaneous Settings Device Selection Select...

Страница 5: ...Short 1 2 DC supply JP17 Short 2 3 DC supply LDO JP19 Short 2 3 DC supply LDO JP28 Short 2 3 DC supply LDO JP29 Short 2 3 DC supply LDO JP26 Open DC supply for ext buffer JP27 Open DC supply for ext...

Страница 6: ...the TSW1200 set up and operation This document assumes the TSW1200 software is installed and functioning properly The front panel of the TSW1200 is shown in Figure 4 The following configuration needs...

Страница 7: ...erator for this input to ensure proper device evaluation Connect input signal through an appropriate BPF at either J6 or J3 Channel A or B i e 170MHz Adjust frequency of generator to match the coheren...

Страница 8: ...input are shown in Table 3 Table 3 Jumper Setting for Transformer Coupled or OPA Driven Input Jumpers or 0 Transformer Coupled Input default OPA Driven Input R119 Install Do not install R123 Install D...

Страница 9: ...ault configuration see CDCE72010 data sheet SCAS858 for details about device default configuration With a 10MHz primary reference at J19 and a 983 04 MHz VCXO on board the CDC outputs a LVCMOS clock a...

Страница 10: ...ct This notice contains important safety information about temperatures and voltages For additional information on TI s environmental and or safety programs please contact the TI application engineer...

Страница 11: ...orized for use in safety critical applications such as life support where a failure of the TI product would reasonably be expected to cause severe personal injury or death unless officers of the parti...

Отзывы: