Weigh Scale Mode

Obtaining a calibration weight:

Before the ADS1232REF can display the mass of an object, it must measure the output of the load cell for
a previously known mass. The known mass can be adjusted.

At power-up, the ADS1232REF expects a 5kg mass. If this mass is not available, or if the load cell range
is not compatible with this mass, the calibration mass can be changed in the following manner:

1. Determine the mass of the calibration weight. The most accurate way to determine the mass is to

weigh the calibration object on an accurately calibrated scale. If the weight is precalibrated, its given
mass can be used, although this approach is not generally as accurate. (Note that the accuracy of the
ADS1232REF as a scale directly depends on the accuracy to which the calibration weight can be
measured.)

2. Switch the mode switch to SCALE mode.
3. Press the PARM buttons simultaneously. The ADS1232REF enters configuration mode.
4. Use the ENT and SEL buttons to select the screen which shows

UNIT=

. This screen allows you to

select the units used for the calibration mass. If the units shown are not correct, adjust them using the
PARM buttons.

5. Use the VALUE buttons to select the screen that shows

CW=

followed by the calibration mass. On this

screen, you can adjust each digit of the calibration mass separately. Select the digit using the VALUE
buttons. The currently-selected digit flashes, and can be adjusted using the PARM buttons.

6. Adjust the calibration mass to match the mass of the calibration weight.
7. Press the PARM buttons to exit the configuration mode.

Preparing Scale mode:

Do the following steps to set up the scale mode:

1. Connect the load cell.
2. If the input shorting jumpers J8 and J7 are connected, disconnect them (see

Figure 1

).

3. Obtain a weight of known mass.
4. If the mass of the weight is not equal to the configured calibration mass, adjust the calibration mass

as

described above.

5. Move the mode switch to SCALE position. If calibration has not been performed, the display reads

NO CAL

.

6. Press the CAL button, and follow the calibration procedure given in

Section 2.3

.

If the calibration is performed properly, and the load cell is connected correctly, the ADS1232REF will
measure the mass of an object placed on the load cell (provided that the mass of the object is within the
device range).

Weigh scale mode is described in detail in

Section 2

.

2

Weigh Scale Mode

In weigh scale mode, the ADS1232REF displays mass.

(1)

Mass is displayed in either SI or avoirdupois

units based on the voltage received from a load cell.

The ADS1232REF operates in Scale mode when the mode switch is set to the Scale position.

To accurately calculate mass, the ADS1232REF must have calibration information for the load cell. When
scale mode is first entered, the ADS1232REF displays

NO CAL

, because the ADS1232REF has no

calibration data when it is powered on.

By default, mass is calculated from the average of four successive readings from the ADC. The number of
points for averaging can be adjusted, and averaging can be turned off.

(1)

Load cells do not measure mass directly; they output a voltage proportional to the weight of an object. The mass of an object can be
accurately inferred from this voltage as long as a calibration is accurately performed, the downward force of gravity remains constant (as
it does if the load cell is not moved to a different altitude), and the tilt of the load cell does not change; if either of the latter conditions
change, a new calibration must be performed.

7

SBAU120B

November 2005

Revised July 2011

ADS1232REF User

'

s Guide

Submit Documentation Feedback

Copyright

©

2005

2011, Texas Instruments Incorporated

Содержание ADS1232REF

Страница 1: ...emote control Designed for very low power consumption Battery 9V or wall power Version 1 1 0 of the firmware includes the following features Weigh scale mode with two point calibration Complete config...

Страница 2: ...REF Hardware Block Diagram 20 9 ADS1232REF PCB Top Side 25 10 ADS1232REF PCB Bottom Side 25 11 ADS1232REF Schematic ADC 26 12 ADS1232REF Schematic MCU 27 13 ADS1232REF Schematic USB 28 List of Tables...

Страница 3: ...odes are taken directly from the ADS1232 device and various measurements are made upon them Several measurements are available including raw display voltage RMS noise and peak to peak measurements The...

Страница 4: ...be set to JTAG 1 3 Powering the Board To apply power to the ADS1232REF connect a 9V battery or plug in a 6V to 9V ac wall adapter AC adapters must be tip positive sleeve negative When an ac adapter is...

Страница 5: ...a 4 wire load cell to the header In this configuration the load cell is excited by the 5V power supply and the ADC reference is taken from the power supply Figure 4 4 Wire Load Cell to Header For thi...

Страница 6: ...e product data sheet are taken with the inputs shorted to 2 5V These noise measurements can be replicated on the ADS1232REF with no external hardware To set up these measurements on the ADS1232REF per...

Страница 7: ...ected disconnect them see Figure 1 3 Obtain a weight of known mass 4 If the mass of the weight is not equal to the configured calibration mass adjust the calibration mass as described above 5 Move the...

Страница 8: ...llowed by st followed by pounds displayed to two decimal places 2 1 1 Calculation of Mass Mass is calculated from ADC code using the formula w m c wzs wt where w mass c the ADC code wt tare weight m w...

Страница 9: ...ns simultaneously The four buttons then assume the functions shown in the CONFIG box To exit Configuration mode press the PARM buttons simultaneously again This function does not cause parameters to b...

Страница 10: ...sted with the PARM buttons The VALUE buttons are used to select the digit The unit is adjusted in the same manner and flashes when selected When the unit is changed the value changes to the equivalent...

Страница 11: ...able units Not all units are available in all modes DISP When this switch is pressed the display shows the current measurement mode and unit While DISP is still pressed pressing UNIT cycles through th...

Страница 12: ...calculation is selected in Configuration mode 50 codes are used in laboratory characterization so this value is the default Peak to peak noise In this mode a number of codes are collected and the abso...

Страница 13: ...ns The choices available are 2 4 8 10 16 32 50 64 100 and 128 The default setting is 50 Voltage reference To convert voltages to codes the ADS1232REF requires the voltage reference level Since this le...

Страница 14: ...talls it and quits You must run the installer again to complete the installation If the Virtual COM Port driver is already installed the installer offers to uninstall it Do not uninstall it cancel thi...

Страница 15: ...ADS1232REF connected The program uses the first ADS1232REF it finds Figure 6 ADS1232REF Software Display Although this procedure is conceptually simple it may not go as smoothly as expected The follo...

Страница 16: ...led and dimmed while Analysis occurs 4 6 1 Averaging The PC software can process data collected from the board using a sliding window averager The controls for the averager are found in the Averager b...

Страница 17: ...normally taken from the 5V power supply If a different reference voltage level is used the value of Vref should be changed to reflect the different level so that voltage calculations are performed co...

Страница 18: ...Vref control it is therefore important that this value be correct 5 Serial Console The ADS1232REF provides a console mode that can be used with any Windows terminal emulation program such as Hyperterm...

Страница 19: ...gain to 64 P3 sets gain to 128 P with no argument prints the current gain setting 5 2 2 R Set Data Rate This command sets the speed of the ADS1232 according to these parameters RF sets rate to fast RS...

Страница 20: ...es the current channel setting 5 2 8 Q Query Parameters Q causes the ADS1232REF to issue a coded string summarizing the current settings The format of the string is P 0 3 R F S OIC 0 1 T followed by a...

Страница 21: ...additional RF rejection 6 4 2 Load Cell Header The load cell header J6 provides a convenient terminal for load cells having a properly fitted header connector It provides excitation and sense connecti...

Страница 22: ...ade from R25 and R26 dividing the power supply by 2 This voltage divider electrically resembles an ideal bridge sensor 6 5 USB Interface The USB interface can be used for firmware download or data com...

Страница 23: ...sistor Panasonic ERJ 3GEYJ152V R4 R10 R16 10K 1 10W 5 Chip Resistor Panasonic ERJ 3GEYJ103V R25 R26 20K 1 16W 1 Chip Resistor Panasonic ERJ 3EKF2002V R18 R27 47K 1 10W 5 Chip Resistor Panasonic ERJ 3G...

Страница 24: ...J8 1 X 2 Header Samtec TSW 102 07 G S J4 J5 3 5mm PCB Terminal Block On Shore Technology ED555 4DS 4 position N A ADS1232 REF PCB Texas Instruments 6474748 D1 30V 200mA Schottky Diode Fairchild Semico...

Страница 25: ...PCB Layout Figure 9 ADS1232REF PCB Top Side Figure 10 ADS1232REF PCB Bottom Side 25 SBAU120B November 2005 Revised July 2011 Schematic and Layout Submit Documentation Feedback Copyright 2005 2011 Texa...

Страница 26: ...DWN 22 SCLK 23 DRDY DOUT 24 U6 ADS1232IPW TP15 GND J5 OST_ED555 4DS J4 OST_ED555 4DS 1 2 3 4 5 6 J6 HEADER 6 5VA EXC EXCSNS SIG SIG EXCSNS EXC NO 1 GND 3 COM 2 V 5 IN 4 U7 TS5A3166DCK 5VA EXC C4 100uF...

Страница 27: ...C2 60 DVSS2 61 U5 M SP430F449IPZ R19 221K R20 221K R21 221K C13 0 01uF C14 0 01uF C15 0 01uF 3 3VD R22 200K SW2 RANGE SW3 TARE SW4 CAL SW5 CONFIG 3 3VD R7 10K R8 10K R9 10K R10 10K R12 100 R13 100 C25...

Страница 28: ...USB RESET TP2 3 3V_USB RX TX RTS DTR C3 4 7uF R17 20 FB SENSE 1 RESET 2 EN 3 GND 4 IN 5 IN 6 OUT 7 OUT 8 U3 TPS77133DGK 2 4 5 3 U10 SN74LVC1G04DBV 2 4 5 3 U11 SN74LVC1G04DBV USB3 3V USB3 3V USB3 3V US...

Страница 29: ...guration section 9 Updated Table 2 9 Revised Section 2 4 1 10 Corrected button nomenclature in Section 3 11 Revised button nomenclature in Section 3 4 13 Corrected button nomenclature errors in Sectio...

Страница 30: ...orized for use in safety critical applications such as life support where a failure of the TI product would reasonably be expected to cause severe personal injury or death unless officers of the parti...

Отзывы: