The Service Monitor program
31
The SD card errors and maintenance messages are not displayed.
Example: Changing the measurement interval in
the Service Monitor for Testomat
®
Modul CL
1st option: Directly on the device via the USB interface
Switch the device to selection mode (see
on page 22).
Connect the computer with the USB socket of the device.
Start the “TestomatModul.exe” program on the computer.
Start the “USB communication” function with the
Manual
and
Alarm
on page 22).
Select the USB port in the service monitor and press the
“Open”
button.
Select the appropriate device under “Device selection”.
Load the firmware version into the service monitor using the
“Read
version” button.
Select “Device” as a data target under the “Import” field and press
the “Import Data” button.
The data from the device is displayed under “Settings”.
Set the desired “measurement interval” with the slider.
Select “Device” as a data target under the “Export” field and re-
s
tore the data with the “Export Data” button.
Close the service monitor with the
“Close” button.
Disconnect the USB connection.
Exit the selection mode with the RESET key.
2nd option: Using the SD card on the computer
Remove the SD card from the device and insert it into the computer.
Start the “TestomatModul.exe” program.
Select “SD card” as a data target under the “Import” field and press
the
“Import Data” button.
The data fro
m the device is displayed under “Settings”.
Set the desired “measurement interval” with the slider.
Select “SD card” as a data target under the “Export” field and re-
s
tore the data with the “Export Data” button.
Remove the SD card from the computer and insert it into the de-
vice again.
Switch the device to selection mode (see
on page 22).
Import the data in selection mode from the SD card to the device
(see
Importing basic programming data from the SD card into the
on page 23)
Exit selection mode with the RESET key.