Adjusting parameters in the device
23
Exiting selection mode
Press the RESET key to exit selection mode.
How to proceed after reset
The software is loaded again after the reset, and the device car-
ries out ventilation prior to analysis.
Start of measurements
Once you have completed the steps in the
you can supply the device with power.
Switch on the device.
In interval operation (continuous operation) the measurement starts
automatically after 15 seconds.
In measurement phase mode, the first measurement is started after a
signal on the stop/start input.
Other functions in selection mode
Importing basic programming data from the SD card into the de-
vice
Press the
Manual
function key.
Press the
100%
function key.
Press the BOOT key.
The yellow LED next to the SD card lights up briefly.
The device then returns to normal interval operation.
Exporting basic programming data from the device onto the SD
card
Press the
Alarm
function key.
Press the
100%
function key.
Press the BOOT key.
The red LED next to the SD card lights up briefly.
The device then returns to normal interval operation.
Resetting to factory settings
Press the
100%
key.
Press the BOOT key.
The device then returns to normal interval operation.
press
press
press
press
press
press
press
press