2 / 24
Handling instructions
•
Repeated switching on/off:
•
Wait at least 5 seconds before switching the unit repeatedly either on or off at the main switch.
•
Observance of the ambient conditions:
•
In order to guarantee reliable operation, only operate the unit under the ambient conditions described in the
technical data section. Always protect the unit against moisture and humidity. It should never come into contact
with condensation or splash water.
•
Safety seal:
•
The original seals attached during manufacture (e.g. EPROM labels) must not be broken, otherwise all warranty
rights are void.
•
Malfunctioning / Repairing a defective unit:
•
The repair of a defective unit – irrespective of the warranty period - is only possible when the unit is dismantled
and returned to us with a description of the error. Please also inform us of the measured medium and the indicator
type. The warranty shall not apply if the controller is tampered with, altered, modified or repaired contrary to the
instructions described in this manual.
•
Before you return the unit for repair work, remove the reagent bottles and ensure that the measuring chamber has
been flushed out and is empty.
•
Electrical load capacity
•
The maximum electrical load capacity of the relay outputs and the total power rating should never be exceeded.
•
Only use the Testomat
®
2000 SO
3
2-
for its intended purpose.
•
Environmental protection regulations
•
Please observe environmental protection regulations and collect any unused reagents and send them to us for
safe disposal in accordance with local statutory requirements.
Operational reliability instructions
Careful handling of the unit increases both its operational reliability and service life!
Therefore, carry out a visual inspection at regular intervals as described below.
•
Are the hose connections of the dosing pump free of leaks?
•
Is there any air inside the dosing hoses?
•
Are all the water connections free of leaks?
•
Are the doors of the unit closed properly?
•
Is the unit unduly soiled?
Maintenance and servicing instructions
(see chapter entitled
Maintenance
)
Safety instructions
•
The unit must be installed and operated in compliance with relevant standards (e.g. DIN, VDE, UVV) or in
accordance with regulations laid down by the respective country.
•
Some functions (e.g. diagnosis, manual operation) allow direct manipulation of the plant to be monitored without
locking or monitoring. Only trained staff should use these functions, which can only be accessed after entering a
password.
•
If you notice that the device is malfunctioning, switch it off immediately. Subsequently shut off the water supply
and contact our service staff.
•
Do not try to repair the unit (loss of warranty rights); always contact authorised service staff. This is the only way to
ensure reliable and safe operation of the plant.
•
After a protective circuit (fuse) has been tripped, attempt to eliminate the cause of malfunctioning (e.g. replace a
defective valve) before reactivating the protective circuit. Frequent tripping is always due to an error which, in
certain circumstances, may also cause damage to the unit.
•
Always observe the safety instructions about working with reagents, chemicals and cleaning agents.
•
Always adhere to hazard warnings and safety tips when using reagents, chemicals and cleaning agents. Please
adhere to the respective safety data sheet! Download the safety data sheets for the supplied reagents at
http://www.heyl.de .
Non-compliance with these instructions can damage the unit as well as the plant and may result in a loss of
warranty rights.