background image

- 12 -

date and time, calibration data and zero offsets. The columns are time (in 

seconds), diffusion stage signal (in fA), 

fi

 lter stage signal (in fA), temperature (in °C), charging current (in nA), 

corona voltage (in kV), 

fl

 ow (in lpm), battery voltage (in V) and status (-) respectively.

Tab-delimited text 

fi

 les can easily be imported into all common data processing programs. 

Checking the Instrument

When you turn on the instrument, it will warm up for 5 minutes. After this interval, it will automatically mea-
sure the zero offsets and start its pump. 
When it is cold, the 5-minute warmup interval will not be enough, and you should let the instrument warm 
up longer to reach a stable temperature. 
Press the menu button once to switch to the raw data screen. You will see the current on each of the stages 
(a few fA). These currents should be reasonably stable. 
Go to the pump screen, and turn the pump off and on – you will hear whether this is working. When the 
pump is turned off, go to the raw data screen, and verify that all stage signals are very close to 0fA, i.e. that 
their absolute values are all smaller than 2fA. To verify proper functioning of the pump and that the instrument 
is sealing, plug the inlet with a 

fi

 nger for a second or two - the pump should increase its speed audibly. If this 

is the case, the instrument may working properly. 

Using the Battery Charger

Please use only the supplied battery charger Mascot.

Battery charger

Fire danger! Never use a power adapter for the testo DiSCmini. 
Only use the battery charger delivered with the testo DiSCmini instrument. 

Charging the battery takes between 2 and 4 hours, depending on the current status of the battery. The LED 
on the charger indicates if it is still charging or if charging is 

fi

 nished. 

Measuring with testo DiSCmini

The purpose of this application note is to ensure that the data you collect with your testo DiSCmini is as 
accurate as possible.
• Turn on the instrument and wait for it to warm up (300 seconds).
• If accurate timing is important, set the instrument clock.
• Unscrew the impactor cap from the impactor body (which can remain in the instrument), and wipe the 

cylindrical impaction surface with a soft paper to remove built up dirt. 

• Verify that charger, pump and stage heater are on (

,  , H visible in display upper right).

• After the warmup phase, the instrument starts measuring, but is not collecting data yet. Optionally per-

form a zero check by connecting a HEPA filter (Article number 98052) to the instrument. Switch to the 
current display (press the right arrow button) and verify that the displayed currents are below 1fA 
absolute.

• Press the recording button once to start recording data. Warning: if you keep it pressed for more than 

about 3 seconds, it will stop recording again. A blinking circle appears on the upper right in the display to 
indicate that the instrument is recording. Keep the instrument running for about 1 minute with the HEPA 
filter, then disconnect it and perform your measurement.

• Reconnect the HEPA filter for another minute after your measurement. Bracketing your measurement 

with zero data is good for quality assurance.

• Press the recording button for about 3 seconds until a beep confirms that the recording has been stopped.
• Run the Java data analysis tool and load the file you just saved. 

If any warnings are displayed, follow the instructions in the application note on software warnings 

Содержание DiSCmini

Страница 1: ...testo DiSCmini Diffusion Size Classifi er Miniature Instruction Manual Version V1 12 September 2016...

Страница 2: ......

Страница 3: ...icators and Connections 7 Aerosol Inlet 7 LCD screen 8 Main Screen 8 Raw Data Screen Main 8 Surface Area Screen Main 9 Advanced mode navigation 9 Charger Screen advanced 9 Flow Screen advanced 9 Heati...

Страница 4: ...nd spare parts purchase ordering contact your local distributor or testo techsup port Email support nanoparticle testo de Sales support and customer service For commercial request please contact your...

Страница 5: ...n not carried out by Testo SE Co KGaA or an authorised service partner use of not genuine spare parts Liability to content The content of this manual is generated with most accurateness Testo SE Co KG...

Страница 6: ...Warnings Warnings for testo DiSCmini Caution Caution means be careful If do not follow the manual instruction you might result an in strument or accessories damage but no human injury Also Caution ref...

Страница 7: ...y damage the instrument High particle concentrations To avoid premature aging of the instrument do not expose it to extremely high aerosol concentrations such as undiluted cigarette smoke Large partic...

Страница 8: ...les will preferably deposit on it by diffusion and detected as an electrical current The remaining particles end up in a second stage the filter stage and are also detected as an electrical current Th...

Страница 9: ...let The aerosol inlet is at the bottom of the instrument The impactor is used to remove larger particles 0 7 microns which would alter the measurements The impactor must always be fitted before starti...

Страница 10: ...or heating and a blinking circle when recording On the lower right corner the remaining battery charge is displayed This is only an approximate value Pressing the right arrow button will lead you to t...

Страница 11: ...advanced menu described below press 3 sec the menu button Press on menu to jump from one screen to the other Charger Screen advanced 3244V 9nA H OFF HIGHVOLT ON 3244V 9nA H OFF HIGHVOLT ON The charger...

Страница 12: ...g table during the offset compensation Date Time Screen advanced SET DATE TIME 2000 04 18 29 SET DATE TIME 2000 04 18 29 You can set the clock and date in this screen Every element year month day will...

Страница 13: ...s the record REC button The testo DiSCmini will indicate that it is recording by blinking a recording symbol in the upper right corner of its display Stop the recording by pressing the REC for few sec...

Страница 14: ...een 2 and 4 hours depending on the current status of the battery The LED on the charger indicates if it is still charging or if charging is finished Measuring with testo DiSCmini The purpose of this a...

Страница 15: ...ts pump once an hour for one minute and continues recording data testo DiSCmini does not use this data immediately instead the offline testo DiSCmini software detects the intervals where the pump was...

Страница 16: ...ameter displays the average particle diameter over time in the graph panel LDSA displays the lung deposited surface area over time in the graph panel Induction correction Rapidly varying charge concen...

Страница 17: ...Temperature displays the instrument temperature over time in the graph panel Flow displays the flow in the instrument over time in the graph panel Note that this is not actual flow in l min but the s...

Страница 18: ...d in the instrument or if the SD card is not formatted properly If this warning appears despite that there is an SD card in the instrument then you should reformat the SD Card File system error If thi...

Страница 19: ...act Testo or your local distri butor for service Noise levels too high Run testo DiSCmini with filtered and dried air for several hours to remove a water film on the insulators which can form at high...

Страница 20: ...2E3 1E6 pt ccm 100nm 5E2 5E5 pt ccm Accuracy 30 in size and number typical Flow rate 1 0 L min 0 1 L min Operating conditions Pressure 800 1100 mbar abs ambient p max at inlet 20mbar Temperature 10 30...

Страница 21: ......

Страница 22: ...0970 3700 en 02 Testo SE Co KGaA Testo Stra e 1 79853 Lenzkirch Deutschland Tel 49 7653 681 5062 Fax 49 7653 681 95062 Web www testo particle com E Mail sales nanoparticle testo de...