testo 6600
- 3.3 Using the software
139
3.3.4.2 2-point
adjustment
Also see
volume 1, chapter 1.3.5.3, page 57
.
1
Expose the reference measuring instrument and the instrument to be
adjusted to the same constant conditions and wait for equalization
period to lapse.
2
Mark the instrument file of the connected instrument.
3
Click on
[Adjusting the transmitter]
.
The
Properties of <Instrument type> <Serial number>
dialogue is
opened with the
Adjusting the transmitter
register.
4
Mark
11.3 %
, enter the reference value for the lower adjustment point
and click on
[Lower adjustment point]
.
The adjustment is carried out.
5
Mark
75.3 %
, enter the reference value for the upper adjustment point
and click on
[Upper adjustment point]
.
The adjustment is carried out.