testo 6651 Ethernet
- 1.4 Operation
65
1.4.6 The individual main menus
1.4.6.1
Editing Main Menu Channel 1
An overview is given in
Overview of the testo 6651 user
menu
(see
chapter
1.4.5, page 63)
.
You can perform basic settings for channel 1.
1 In Measuring Mode, press Set, select Main Menu Channel 1 using
or
and confirm selection with set.
One of the following parameters can now be selected using
or
, after
which the selection must be confirmed with SET:
Channel 1 unit
The parameter for this channel is selected.
Selection: % RH, °C, °F, °Ctd, °Ftd,
g/m
3
.
Edit/select parameter with
or
, confirm with SET or abort input
with ESC.
Scale minimum for channel 1
The lower scale limit is edited;
Unit as selected above (example: 4 mA = 0 % RH).
Editing the value: Scroll one digit to the right using
, increase value
of digit by 1 using
. Confirm with SET or abort entry with ESC.
Scale maximum for channel 1
The upper scale limit is edited;
Unit as selected above (example: 20 mA = 100 % RH).
Editing the value: Scroll one digit to the right using
, increase value
of digit by 1 using
. Confirm with SET or abort entry with ESC.
Signal delay ("Damping")for channel 1
The analog signal can be delayed ("Damping"); a time constant is
selected for this (1 = no delay; 15 = longest delay).
Edit/select parameter using
or
, confirm with SET or cancel entry
via ESC.
2 Continue to Main Menu Channel 2 using
or
or return to
Measuring Mode by pressing esc.
1.4.6.2
Editing main menu of channel 2
See channel 1.