background image

Changing the batteries

Testo 550 uses 4x 1.5 V, AA batteries.To 
replace the batteries please follow these few 
steps below:

1.  Fold out the hook (Pic. 1).

2.  Grab the clip and squeeze it together 

and remove the cap (Pic. 2).

3.  Insert/Change the batteries.  

Observe the polarity.

Power ON / OFF

1.  Connect probes to the testo 550 prior 

to powering it up.

2.  Press the power button [   ] to turn testo 

550 on.

3.  All display segments are lit (2 s.)
4.  Measurement view is displayed
5.  Press the power button to turn the testo 

550 off.

Set the refrigerant

1.  Press the [R, Start/Stop] button so you 

can choose the required refrigerant 
(Pic. 3).

2.  Use the arrow keys [

pq

] to scroll 

through the choices.

3.  Press the [R, Start/Stop] button to set 

the chosen refrigerant.

Set the Units / Altitude / Mode

1.  Press the [Set] button once to get to 

temperature units menu (Pic. 4).

2.  Choose the required units with the  

arrow keys [

pq

].

3.  Press the [Set] button for the second 

time to be able to chose the pressure 
units (Pic. 5).

4.  Choose the desired units with the arrow 

keys.

5.  Press the [Set] button for the third time 

so you select absolute or relative  
pressure (Pic. 6).

6.  Press the [Set] button four times, so 

you can set the barometic pressure in 
inHg (Pic. 7) .

7. 

Press the [Set] button five times to 

select the AC/R [ ], heat pump [ ], or 
Auto [

]. (Pic 8).  Auto mode senses if 

the Low pressure side is 15 psi higher 
than the High pressure side, and will 
automatically reserve the display of the 
High / Low side pressures

8.  Press the [Set] button six times to exit 

the [Set]-menu.

Backlight

Testo 550 has a backlight to 
improve viewing in dark or daylight 
conditions. Press the backlight 
[

k

] button to turn the backlight 

on. Press it once again to turn the 
backlight off.

Pressure zeroing

Please zero the pressure sensors before every use of the 
testo 550.

1.  Loosen the hose connections

2. 

Open the valve knobs to confirm no pressure in the 

manifold

3.  Press the [p=0] button

The sensors are now zeroed and ready for measurement.

Superheat / Subcooling

The testo 550 calculates  superheat and subcooling in 
real time.
1.  Connect the temperature probes to testo 550.
2.  Connect testo 550 and the probes to the air condi-

tioning, refrigeration or heat pump system.

3.  Switch on testo 550.
4.  You will now see the calculated  

evaporation and condensation 
and temperature (Ev and Co) and system pressures 
at the bottom of the display, (Pic. 8).

5.  Press the UP-arrow [

p

] once to see the tempera-

ture difference (

D

) (Pic. 9).

6.  Press the UP-arrow [

p

] twice to see the real time 

superheating and 
subcooling (SH and SC) (Pic. 10).

7.  Press the UP-arrow [

p

] for the third time to see the 

real time measured temperature (T1 and T2) (Pic. 
11).

8.  Press the UP-arrow again to get to back to the 

calculated evaporation and condensation tempera-
ture.You can also use the DOWN-arrow to switch 
between the menus but the order will be reversed.

Leak Test

The temperature compensated leak test is used to indi-
cate leaks in an AC/R or heat pump system.
1.  Plug in the temperature probe into the high side of the 

instrument.

2.  With the instrument off, simultaneously press the Set 

and Mode while turning on the 550. This will turn off the 
surface compensation to provide an accurate ambient 
air temperature. Temperature surface compensation is 
turned back on the next time that the 550 is turned on.

3.  Zero the pressure sensors.
4.  Connect testo 550 to the system.
5.  Press the [Mode] button once to get to the leak test 

mode (Pic. 12).

6.  Now press the [R, Start/Stop] button to start the test.
7.  Press arrow keys [

p

,

q

] to see the  

measured temperature (Pic. 13).

8.  The test duration depends on the system size.
9.  Press [R, Start/Stop] again to stop the test.
10.  The results will be displayed.
11.  Press the [Mode] button twice to return to the normal 

measurement mode.

Note:

 

r

P could be different from starting 

and final test pressure 

r

P as the true 

r

p is 

calculated from the gas laws.

Pic. 13

Pic. 12

Pic. 11

Pic. 10

Pic. 9

Pic. 8

Pic. 7

Pic. 6

Pic. 5

Pic. 4

Pic. 3

Pic. 2

Pic. 1

Pic. 8

Отзывы: