Operation

Switch-on

Switch detector on in fresh air since the sensor is initialised when 
switched on

Readiness to operate is indicated by an audible signal.

Detecing leaks 

  Position the detector such that the sensor opening (orange) is 

purged by the ambient air in the area under scrutiny. Avoid 
covering the surface of the sensor.

-  The 8 bar symbols change to black according to the concen-

trations detected and the frequency of the audible signal rises 
with increasing gas concentrations.

-  If the alarm value is exceeded, a continuous tone is emitted 

and the alarm symbol ( ) lights up.

Switch-off

Maintenance and Service

Changing the battery

1  Open battery section
2  Remove spent batteries 

and replace with new ones. 
O/-.

3  Close battery section.

Cleaning the detector

  The detector housing should be cleaned with a damp cloth 

(soap water) if dirty. Do not use abrasive cleaning agents or 
solvents.

Troubleshooting

Problem 

 

Possible causes 

Remedy

 lights up in display. 

Battery life < 15min. 

  Change batteries.

Detector cannot be  

Batteries are spent. 

  Change batteries. 

switched on.
Error lights up in display. 

Detector lay for a longer  

  Clean sensor by 

   

 

period of time in heavily  

  repeatedly switching on 

   

 

contaminated air or the  

  detector, letting it heat up 

   

 

sensor is clogged with  

  and switching it off again 

  

 

dirt. 

 

   If the Error message still  

does not disappear after  

detector has been 

switched on several times, 

switch it off and contact 

Testo Customer Service or 

your nearest Testo dealer.

Spare parts

Name 

 

 

 

 

Part  no

Carrying 

loop 

 

    0516 

3181

Carrying loop with belt clip 

 

 

 

0516 0317

General notes

Please read this documentation through carefully and familiarise 
yourself with the operation of the product before putting it to use. 
Keep this document to hand so that you can refer to it when 
necessary. 

Safety Information

Avoid electrical hazards:

  Do not measure on or near live parts. 

Protect instrument:

  Never store instrument together with solvents (e.g. acetone).  

Preserving product safety / warranty claims:

  Operate the instrument only within the parameters specified in 

the Technical data. 

  Operate the instrument properly and according to its intended 

purpose. 

  Do not use force. 
  Open the instrument only when this is expressly described in 

the documentation for maintenance or service purposes. 

Segment test

Detector off

Ready to 

operate

1s

60s

3s

Detector off

Ready to operate

testo 317-2

Gas Detector

Instruction manual 

en

Heating up 

phase/Self-test

Ensure correct disposal:

  Disposal of defective rechargeable batteries and spent batter-

ies at the collection points provided. 

  Send the instrument directly to us at the end of its life cycle. 

We will ensure that it is disposed of in an environmentally 
friendly manner.

Intended Use

testo 317-2 is a practical gas detector which detects methane 
(CH

4

) and propane gas (C

3

H

8

).

Do not use the testo gas detector as a monitoring instrument 
for your own personal safety!

Technical Data

Feature  

Values

Concentration ranges 

0 to 20,000ppm CH4, 0 to 10,000 ppm C3H8

Display 

 

8 segment trend display

Alarm thresholds 

10,000ppm CH4, 20% LEL, 5,000ppm C3H8, 20% LEL

Lower response thresholds 

100ppm CH4, 50ppm C3H8

Response time t90 

<5s

Heating-up time 

60s

Power supply   

2 batteries, type micro AAA 1.5 V (LR03)

Battery life 

 

Approx. 4h

Transport/Storage temperature  -20 to +50°C
Ambient temperature 

-5 to +45°C

Audible signal emitter 

85 dB (A)

EU directive 

 

2014/30/EC

Warranty measuring instrument: 24 months
Warranty sensor: 

12 months