background image

7 Maintaining the product 

 

 

17

 

[

]

 lights up. 

6.4.

 

Bluetooth

®

 and IrDA data transfer 

 

 

 

Data is transferred via Bluetooth

®

 if both instruments have 

this interface. Otherwise, data is transferred via the IrDA 

interface. 

 

 

Data can currently be transferred to the following testo measuring 

instruments: 
•  testo 330 (0632 3306 / 0632 3307) from firmware version V1.11 
•  testo 330 (0632 3304 / 0632 3305) from firmware version V1.63 

  testo 315-3 is on and a measurement has been carried out. 

>  Switch on data transfer: press and hold down 

[ ]

[ ]

 lights up. Data is transferred. 

 

 

 

Please refer to the operating instructions for the relevant 

testo measuring instrument. 

 

  

7

 

Maintaining the product 

 

Charging the rechargeable battery 

1.  Connect the mains unit connector to the instrument's micro 

USB socket. 

2.  Connect the mains plug to the mains socket. 
The charging process will start automatically. Charging is indicated 
by a change in the battery icon segments. 

 lights up when the 

battery is fully charged. 

 

Cleaning the instrument 

Do not use any aggressive cleaning agents or solvents! Mild 

household cleaning agents and soap suds may be used. 

Содержание 315-3

Страница 1: ...testo 315 3 CO CO2 measuring instrument Instruction manual...

Страница 2: ...2...

Страница 3: ...er instrument data 7 3 2 4 Bluetooth module option 7 4 Product description 9 4 1 Overview 9 4 1 1 Control elements and connections 9 4 1 2 Display 12 4 2 Basic properties 13 5 First steps 13 6 Using t...

Страница 4: ...writing standards Represen tation Explanation Note Basic or further information 1 2 Action more steps the sequence must be followed Action a step or an optional step Result of an action Menu Elements...

Страница 5: ...mperatures For products with Bluetooth optional Changes or modifications that have been made without the explicit consent of the responsible approval authority may cause the retraction of the type app...

Страница 6: ...ndustry The Bluetooth option may only be operated in countries in which it is type approved 3 2 Technical data 3 2 1 Measurement ranges and resolution Measurement parameter Measuring range Resolution...

Страница 7: ...to 95 RH Weight incl holster Approx 382 g Housing material ABS Dimensions L x W x H 190 x 65 x 40 mm Protection class IP 40 acc to EN 60529 Operating time Min 10 h measuring time at 20 C 68 F mains o...

Страница 8: ...inland FI France FR Greece GR Ireland IE Italy IT Latvia LV Lithuania LT Luxembourg LU Malta MT Netherlands NL Austria AT Poland PL Portugal PT Romania RO Sweden SE Slovakia SK Slovenia SI Spain ES Cz...

Страница 9: ...4 1 Overview 4 1 1 Control elements and connections 1 Connection for temperature humidity module optional 2 IR interface record printer 3 CO sensor module 4 CO2 sensor module back 5 Display 6 Keyboar...

Страница 10: ...figuration mode Confirms entry next setting Long Reading display current value Opens configuration mode Reading display maximum value minimum value hold value Resets maximum value minimum value hold v...

Страница 11: ...n mode Changes option increases value Long Reading display Sets zero point deletes zero point Configuration mode Increases value quickly Short Reading display Prints data Configuration mode Changes op...

Страница 12: ...abled Lights up information Flashes error message Differential measurement calibration Flashes adjust CO2 sensor in fresh air CO2 sensor should be adjusted in fresh air after 120 operating hours 2 Mea...

Страница 13: ...45 C 32 to 113 F 5 First steps Switching on Connect the humidity temperature module 0636 9725 before switching the instrument on option 1 Switch the instrument on press briefly All segments of the di...

Страница 14: ...cancel the function press briefly until the display changes Adjustable functions Ensure correct settings all settings are transferred immediately There is no Cancel function Settings and display of t...

Страница 15: ...e carried out on busy roads or in closed rooms Ensure that before and during the calibration adjustment no exhaled air reaches the instrument Before completing the adjustment the instrument should be...

Страница 16: ...Hold Max Min testo 315 3 is in measurement view The measuring values for Hold Max Min can be called up for CO and CO2 Temperature and humidity if temperature and humidity module is connected Dew poin...

Страница 17: ...en carried out Switch on data transfer press and hold down lights up Data is transferred Please refer to the operating instructions for the relevant testo measuring instrument 7 Maintaining the produc...

Страница 18: ...and E290 lights up The current CO2 measuring value is too high for an adjustment Acknowledge the error message with Hold Max Min and carry out the adjustment again in fresh air CO2 measuring value 65...

Страница 19: ...ew point measurements be carried out Humidity temperature and dew point measurements are possible with the temperature humidity module accessory The display is illuminated also after the testo 315 3 i...

Страница 20: ...ts Description Article no testo 315 3 without Bluetooth incl micro USB mains unit cable USB A USB micro B 0632 3153 testo 315 3 with Bluetooth incl micro USB mains unit cable USB A USB micro B 0632 31...