6 First steps 

 

6  First steps 

 

6.1 

Inserting/changing batteries 



The instrument is switched off. 

 

Open the battery compartment (on the back of the instrument) via the 
snap lock. 

 

Insert batteries/rechargeable batteries (pay attention to the polarity!). 

 

 

Close the battery compartment. 

 

 

When not in use for a long period: Take out the (rechargeable) 
batteries. 

 

 

6.2 

Getting to know the product 

 

6.2.1 

Switching the instrument on and off 

Switching on 

Only switch on the instrument in fresh air, since automatic zeroing is carried out 
when the instrument is switched on. The ambient temperature and ambient 
humidity during zeroing should correspond to the ambient conditions at the 
measuring location. If necessary, zero again manually at the measuring location 
(switch off and on again). 

 

If the unit is not used for a prolonged period of time, the sensor will 
become contaminated. Particularly if the instrument has not been in 
operation for a prolonged period of time (> 2 weeks), it should be left 
switched on for a while before being used. The longer it has not been in 
operation, the longer this additional warming-up phase should be. 
Please note that the instrument switches itself off by default after 10 
min of inactivity. 

 

Press and hold down (1 sec) the 

On/Off

 key. 

 

 

Warm-up phase (HEAT) 

The instrument starts up. With regular use, the warm-up period takes 
approx. 20 sec and is symbolized by the flashing and emptying bar 
display. 
As long as the sensor LED is flashing orange, the instrument is not 
ready for use. 

 

Содержание 0560 3161

Страница 1: ...testo 316i gas leak detector 0560 3161 Instruction manual...

Страница 2: ......

Страница 3: ...7 6 3 Establishing a Bluetooth connection 8 6 3 1 Establishing a Bluetooth connection to the testo Smart App 9 7 Using the product 9 7 1 Controls 9 7 1 1 Implementing settings 11 7 2 Carrying out a fu...

Страница 4: ......

Страница 5: ...he technical data Do not apply any force Do not operate the instrument if there are signs of damage to the housing or connected cables Dangers may also arise from objects to be measured or the measuri...

Страница 6: ...ormation denoted by the following warnings Implement the precautionary measures specified DANGER Risk of death WARNING Indicates possible serious injury CAUTION Indicates possible minor injury CAUTION...

Страница 7: ...ot allow the sensor to come into contact with moisture or acids as it will react cross sensitively 4 Intended use The testo 316i is a gas leak detector for the short term detection of leaks in gas sys...

Страница 8: ...ED 3 Flexible sensor tube 4 Visualization of gas concentration 5 Control keys 6 Battery compartment on rear Symbol explanation Do not allow children under 6 years of age to play with batteries Do not...

Страница 9: ...ient conditions at the measuring location If necessary zero again manually at the measuring location switch off and on again If the unit is not used for a prolonged period of time the sensor will beco...

Страница 10: ...tion above the warning threshold the instrument switches itself off The switch off is signalled beforehand by an alarm sound red flashing of the sensor LED and the bar display top element lights up re...

Страница 11: ...th devices in the vicinity 2 In the Bluetooth menu check whether the required instrument is connected If necessary switch the instrument to be connected off and on again to restart the connection mode...

Страница 12: ...ser interface of the app not numbered One level back Exit view Share measurement data report Searching Favourite Deleting Further information Display report Multiple selection The Main menu can be acc...

Страница 13: ...setting functions 1 Press the relevant key to select the functions Secondary assignment long press All keys with a white corner have a secondary assignment which can be selected by pressing and holdin...

Страница 14: ...ement parameter Cycle through CH4 methane C3H8 propane C4H10 butane or H2 hydrogen Sensor LED Switch the LED on the sensor head off or on Alarm sound ON acoustic alarm on or OFF acoustic alarm off The...

Страница 15: ...ATTENTION Destruction of the sensor due to external influences Do not expose the sensor to high concentrations of H2S hydrogen sulphide SOx sulphur dioxides Cl2 chlorine or HCl hydrogen chloride Preve...

Страница 16: ...unit changes to VOL and the bar display increases Concentration 9999 ppm 0 99 VOL Sensor LED and the upper elements of the bar display light up red alarm threshold If the warning threshold 100 ppm is...

Страница 17: ...cloth Do not use any aggressive cleaning agents or solvents Mild household cleaning agents and soap suds may be used Storage transportation To prevent contamination of the sensor please do not store...

Страница 18: ...anently installed sensor head that can only be replaced by Testo Customer Service 9 Technical data for testo 316i Feature Value Measurement parameters via app ppm LEL Detectable gases Methane propane...

Страница 19: ...arm up phase Battery voltage too low Replace batteries 10 1 1 Hard Reset 1 If you encounter any problems with the firmware press and hold the ON OFF key for a long time 4 sec to perform a reset 10 2 A...

Страница 20: ...Testo SE Co KGaA Celsiusstr 1 79822 Titisee Neustadt Germany Phone 49 0 7653 681 0 Email info testo de www testo com 0970 3261 en 01 08 2022...

Отзывы: