TESTO 0560 0110 Скачать руководство пользователя страница 23

8

 

Controls via testo Smart App

 

22 

  Making measuring instrument settings 

 



The measuring instrument is connected to the testo Smart App. 

 

Click on 

 

The main menu opens. 

 

 Click on 

Sensors

 

The 

Sensors

 menu opens. 

 

Click on the required measuring instrument. 

 

Information about the model, order number, serial number and 
firmware version is displayed. 

 

Click on the 

Settings

 tab. 

 

A window with settings for the respective measuring instrument 
opens. 

In addition to the settings that can be performed on the measuring 
instrument, additional settings can also be made. 

 

Содержание 0560 0110

Страница 1: ...testo 110 Temperature meter for NTC and Pt100 0560 0110 Instruction manual...

Страница 2: ...onnection to the testo Smart App 10 Using the product 11 Controls on the measuring instrument 11 Implementing settings 12 Opening configuration mode 13 Setting the unit 13 Setting alarm thresholds 14...

Страница 3: ...trument menu 24 Setting Auto Off 25 Activating damping 25 Configuring alarms 26 Setting the surface increment 26 Display of the readings 27 Adjusting the view 27 Exporting readings 28 Excel CSV Export...

Страница 4: ...the technical data Do not apply any force Do not operate the instrument if there are signs of damage to the housing or connected cables Dangers may also arise from objects to be measured or the measu...

Страница 5: ...nformation denoted by the following warnings Implement the precautionary measures specified DANGER Risk of death WARNING Indicates possible serious injury CAUTION Indicates possible minor injury CAUTI...

Страница 6: ...ing Intended use The testo 110 is a compact measuring instrument for measuring temperatures The product is designed for the following tasks areas Food sector Laboratory sector The product should not b...

Страница 7: ...6 Product description Instrument overview 1 Control keys 2 Display 3 Socket for NTC or Pt100 type probe with TUC connector 4 Speaker for alarm sound 5 Battery compartment Explanation of icons Refer t...

Страница 8: ...nsert or replace batteries 3 x AA batteries Observe the polarity 3 Close the battery compartment When not in use for a long period Take out the rechargeable batteries Symbol explanation Do not allow c...

Страница 9: ...nstrument on and off Switching on 1 Press and hold down 2 sec the On Off key Measurement view is opened The current reading is displayed or lights up if no reading is available Switching off 1 Press a...

Страница 10: ...onnection for the first time press and hold approx 2 sec the key While the instrument is trying to establish a Bluetooth connection the icon flashes on the display Bluetooth remains enabled until it i...

Страница 11: ...ument 1 Open the testo Smart App The app automatically searches for Bluetooth devices in the vicinity 2 In the Sensors menu check whether the required instrument is connected If necessary switch the i...

Страница 12: ...controls are implemented either on the instrument or via the app If the measuring instrument is connected to the testo Smart App settings can only be made via the app The measuring instrument then re...

Страница 13: ...e correct settings all settings are transferred immediately There is no Cancel function Function Setting options comments Bluetooth long press Switch the Bluetooth connection on or off Arrow left Free...

Страница 14: ...hed on and is in the measurement view 1 Press MENU ENTER until the display changes The instrument is now in configuration mode Press MENU ENTER to switch to the next function You can leave configurati...

Страница 15: ...of the setting menus Configuration mode is open M RES menu reset is displayed 1 Press to select the required option and confirm with MENU ENTER o NO Do not carry out reset o YES Carry out reset All c...

Страница 16: ...m and minimum values since the last time the instrument was switched on in the standard view or during a multi point or continuous measurement can be displayed 1 Press several times until the desired...

Страница 17: ...asurement and calculate the mean value Press MODE END and flash The number of measured values and the calculated multi point mean value are displayed 4 To switch back to the measuring view Press MODE...

Страница 18: ...asuring view Press MODE END Printing data A testo Bluetooth IRDA printer order no 0554 0621 is connected via Bluetooth and switched on When first setting up a connection between the testo measuring in...

Страница 19: ...view and settings menus cannot be opened on the measuring instrument Overview of operating controls 1 Choice of applications 2 Switch between the views live graphic table 3 Display of connected measu...

Страница 20: ...nt data report Searching Favourite Deleting Further information Display report Multiple selection The Main menu can be accessed via the icon at top left To exit the main menu select a menu or right cl...

Страница 21: ...you can find the version number of the installed App 1 Click on 2 Select Help and Information 3 Select Instrument information The version number of the app and the ID are displayed Displaying the tuto...

Страница 22: ...Your selected application is displayed Setting favourites 1 Click on A selection of menus for various applications is displayed 2 Select the application you want to set as a favourite 3 Click on The...

Страница 23: ...ensors The Sensors menu opens 3 Click on the required measuring instrument Information about the model order number serial number and firmware version is displayed 4 Click on the Settings tab A window...

Страница 24: ...pen an input window to enter a specific value or select a unit For details on the settings options see the following sub sections Changes to the measuring instrument settings in the app are transferre...

Страница 25: ...tab in the Sensors menu is open 1 Activate Customization instrument menu and click on the blue text Change instrument menu list under the settings heading The Customization instrument menu dialogue op...

Страница 26: ...the menu reset M RES and then all settings menus will be displayed on the measuring instrument again Setting Auto Off The Settings tab is open 1 Enable Activate Auto off using the slider The measuring...

Страница 27: ...ctivating and defining upper and lower warning and alarm values is displayed 5 Click OK to confirm the settings Changes to the measuring instrument settings in the app are transferred directly to the...

Страница 28: ...p to four different readings can be displayed in graph format Readings to be displayed can be selected by tapping on a reading above the diagram Table view In the table view all readings are displayed...

Страница 29: ...8 Controls via testo Smart App 28 Exporting readings Export Report Delete 1 Click on 2 Select Memory 3 Click on Select measurement...

Страница 30: ...reate PDF with all readings button 3 Click on Create For measurements please be aware that the option Create PDF with all readings is only possible up to 30 pages due to the resulting file size and nu...

Страница 31: ...ap lock 2 Insert or replace batteries 3 x AA batteries Observe the polarity 3 Close the battery compartment When not in use for a long period Take out the rechargeable batteries Cleaning the instrumen...

Страница 32: ...according to digital probe Measuring range NTC 50 to 150 C Pt100 200 to 800 C Operating temperature 20 to 50 C Storage temperature 20 to 50 C Operating humidity 0 to 80 RH IP class IP20 Level of conta...

Страница 33: ...ssible measuring range has been exceeded Keep within the permissible measuring range Display UUUU Permissible measuring range has been undershot Keep within the permissible measuring range Display BT...

Страница 34: ...terrupted Keys are locked for 3 s Restart the measuring instrument and testo Smart App and check the Bluetooth connection If we have not been able to answer your question please contact your local dea...

Страница 35: ...nts on pipes 5 65 mm 0615 5605 Stub temperature probe digital with NTC temperature sensor 0572 2162 Stub humidity temperature probe digital 0572 2164 Humidity temperature cable probe digital 0572 2165...

Страница 36: ......

Страница 37: ...Testo SE Co KGaA Celsiusstr 2 79822 Titisee Neustadt Germany Phone 49 0 7653 681 0 E mail info testo de www testo com 0970 9110 en 01 10 2022...

Отзывы: