21
20
Temperature measurement
0.01 °C resolution for Pt100 and thermocouple probe
i
Function buttons
Pt100 probes are generally shown with a resolution of 0.01 °C. This
resolution is possible for a reading to max. +300 °C. At higher
temperatures, the symbol for a disconnected probe appears (- - - - -).
If you wish to measure at over +300 °C you have to switch to a
resolution of 0.1 °C: allocate “0.1...0.01” to function button and
activate.
The switch-over is only possible if a measurement program is not
activated. It must otherwise be cancelled beforehand.
Switch thermocouple probe with EEPROM following adjustment to
0.01 °C resolution.
The following also applies: maximum measured value = +300 °C.
Temperature measurement
Optimum accuracy only applies in the environment of the calibration
point. There may be deviations from the true desired value in other
temperature ranges.
Warning:
If a probe is newly adjusted, the factory calibration
data is deleted. You can retrieve the original values
via
PROBE - R
ESET
.
➁
Measurement:
Place the function display at the top left on the
Save
(= disk symbol) symbol.
Return to the measurement menu by pressing
. Switch off
measuring instrument. Plug in the newly calibrated probe to the
instrument on location and switch on. The probe socket is freely
selectable. Select the
Location
function field via
and set
the required
location
with .
Press the
button after every measurement and select the
required
location
before the next measurement
.
A log is set up in the hand-held instrument memory each time the
button is pressed. The log includes the location, date and time,
all additional information, set correction values... and can be printed
under
PRINT - C
ONFIGURATION
.
➂
Printing
all of the measured data following completion of the
measurement series: set the measuring location at top right. Call up
the log in the
Main menu
under
MEMORY
-
R
EADOUT
and print by
pressing
.
Note: The printer to be used when printing must be selected via
the menu PRINT -
PRINTER TYPE
.
Example:
Log with attachable
printer
Flow -
return