6.8 Adjusting the distance between the mixing bowl and mixer
If the distance between the mixing bowl
and mixer is bigger than the applicable
standard, this can be adjusted.
First, the mixing bowl must be lowered
and removed
(see point 6.3).
Then loosen and remove the mixer (see
point 6.4)
Now the two clamping screws (see Figure
1), on the holder for the mixing bowl can
be loosened by approx. 1 rotation.
Figure 1
Now loosen the lock nut of the adjusting
screw (see Figure 2), at the upper end of
the mixing bowl holder.
Then use the adjusting screw to set the
distance.
Turning the adjusting screw clockwise
will lessen the distance.
Turning the adjusting screw anti-clock-
wise will increase the distance.
For checking, the distance template
(TESTING Art. No.: 1.0203.07) can be
used to check the distance between the
mixer and the bowl.
Figure 2
If the distance is correctly set, the lock nut, the adjusting screw and the two clamping
screws must be firmly tightened on the bracket for the mixing bowl. The contact protec-
tion must be re-assembled, the mixing bowl and the mixer can now be used.
Clamping screw
Clamping screw
Lock nut
Adjusting screw
Содержание 1.0205
Страница 48: ...Laboratory mortar mixer 1 0205 48 22 I O board...
Страница 49: ...Laboratory mortar mixer 1 0205 49 23 Circuit diagram...
Страница 51: ...Laboratory mortar mixer 1 0205 51 24 2 Mixing program 2 according to EN 480 T1...
Страница 52: ...Laboratory mortar mixer 1 0205 52 24 3 Mixing program 3 according to EN 196 T3...
Страница 53: ...Laboratory mortar mixer 1 0205 53 24 4 Mixing program 4 according to EN 196 T9...
Страница 55: ...Laboratory mortar mixer 1 0205 55 25 2 Mixing program 2 according to ASTM C451...
Страница 56: ...Laboratory mortar mixer 1 0205 56 25 3 Mixing program 3 according to ASTM C305 paste with start delay...
Страница 57: ...Laboratory mortar mixer 1 0205 57 25 4 Mixing program 4 according to ASTM C305 paste without start delay...