Chapter 6 – Specifications
TestEquity 3007C Temperature Chamber
Page 6-1
Chapter 6 – Specifications
Model 3007C Chamber Specifications
Temperature Range
–73°C to +175°C
Control Tolerance
±0.5°C (Measured at the control sensor after stabilization)
Uniformity
±1.0°C (Variations throughout the chamber after stabilization)
Live Load Capacity @
+23°C
0°C
–20°C
–40°C
–55°C
–65°C
3700 W 3500 W 3300 W 2700 W 2300 W 1800 W
Cooling Rate, Compressors Only (empty chamber)*
End Temp
Start Temp to
+23°C
0°C
-40°C
-55°C
-65°C
-73°C
+85°C
18°C/min 14°C/min 12°C/min 11°C/min 10°C/min Ultimate
Cooling Rate, Compressors Only (25 lb static aluminum load)*
End Temp
Start Temp to
+23°C
0°C
-40°C
-55°C
-65°C
-73°C
+85°C
14°C/min 10°C/min 8°C/min 7°C/min 6°C/min Ultimate
Cooling Rate, LN2 Boost and Compressors (empty chamber)*
20°C/minute from +85°C to -40°C
Cooling Rate, LN2 Boost and Compressors (25 lb static aluminum load)*
13°C/minute from +85°C to -40°C
Heat Up Ramp Rate (empty chamber)*
17°C/minute from -40°C to +85 °C
Heat Up Ramp Rate (25 lb static aluminum load)*
12°C/minute from -40°C to +85 °C
* NOTE:
Ramp rates are measured at the control sensor after a 2 hour soak at the respective start temperature with
an empty chamber. Measured with setpoint set beyond the start and end temperatures. Ramp rates do not include the
effect of proportional band when approaching setpoint.
Input Power Requirements
230 V ±10%, 60 Hz, 3 PH
Max Current Draw 39 A; Recommended Service 50 A
208 V -5/+10%, 60 Hz, 3 PH
Max Current Draw 35 A; Recommended Service 45 A
Input may be configured for nominal 208 V or 230 V in the field by changing jumpers.
Call for other voltages or 50 Hz operation.
Workspace Dimensions
24" W x 21" H x 24" D (7 cubic feet)
Outside Dimensions
33" W x 79" H x 54.25" D (nominal)
Door latch adds 3" to width on right side, may be removed for move-in.
Min. Installed Clearance
12" from the left and right side, 24" from the rear
Access Ports
4" Port on left and right side (two total), Supplied with foam plugs
Weight
1,200 pounds
Heat of Rejection
27,500 BTUH
NOTE:
Performance is typical and based on operation at 23°C (73°F) ambient and nominal input voltage. This
product is designed for use in a normal conditioned laboratory. Operation at higher ambient temperatures may result
in decreased cooling performance. Additional ports and shelves will also affect performance. Operation above 30°C
(85°F) or below 16°C (60°F) ambient is not recommended.