Chapter 8 – Maintenance
Page 8-4
TestEquity 1007S Temperature Chamber
How to verify the chamber performance.
These tests verify the performance of the heating, refrigeration, electrical controls, temperature
controller, and air circulation systems. The chamber should meet all published performance
specifications if all of these tests are successfully passed.
These tests assume that the Temperature Controller’s setup and tuning values have not been
changed from the values as shipped from TestEquity. Also, the Limit Controller high limit must
be set to over +85°C (+88°C would be fine), and the low limit set to –75°C.
If the chamber fails any of these tests, it should be removed from service to prevent further
damage until the cause of the problem is determined and resolved.
1.
The chamber interior should be empty and at ambient temperature, approxi23°C.
2.
Set the Temperature Controller Set Point to +85°C and turn the Conditioning Switch ON.
3.
The Heat Light should be ON continuously and the Cool Light should be OFF.
4.
The chamber should heat up to about +80°C and begin controlling (Heat Light cycles
ON/OFF) within 7 minutes.
5.
The chamber temperature should slowly increase and stabilize to +85°C. It should NOT
overshoot 85°C by more than a few tenths of a degree, and the compressors should
NOT need to turn ON in order to ma85°C.
6.
After stabilization, the chamber temperature should vary no more than ±0.5°C, or a total of
1°C.
7.
Let the chamber stay at +85°C for two hours.
8.
After two hours at +85°C, set the Temperature Controller Set Point to –30°C.
9.
The compressor should turn ON within a few seconds. The Cool Light should be ON
continuously and the Heat Light should be OFF.
10.
The chamber should cool down to about –25°C and begin controlling (Cool Light cycles
ON/OFF) within 18 minutes.
11.
The chamber temperature should slowly decrease and stabilize to –30°C. It should NOT
undershoot beyond –30°C by more than a few tenths of a degree, and the compressors should
NOT need to turn OFF in order to maintain –30°C.
12.
After stabilization, the chamber temperature should vary no more than ±0.5°C, or a total of
1°C.
13.
Set the Temperature Controller Set Point to +23°C. The chamber should begin to heat up.
The compressors should turn off within approximately 1 minute.
14.
This concludes the chamber performance verification tests.
15.
Let the chamber heat up to +23°C before turning the Conditioning Switch OFF.