4. Installation

a.  Simply plug-in the BNC output connector to the vertical input of a general purposed 

oscilloscope or other measurement instrument, and connects the auxiliary grounding 

terminal to a proper ground. The measurement instrument must have a ground referenced. 

b.  Connect an appropriate power source to this probe and or enter the batteries, then turn it 

on.   

c.  Select the proper attenuation ratio. When measuring signals below 700V switch the 

attenuation ration to 1/100 in order to get higher resolution and less noise ratio. Otherwise, 

set the attenuation ratio to 1/1000. 

                               

WARNING. To protect against electric shock, use only the   

                                                      accessories supplied with this probe. 

                                                               

d. Using the appropriate probe accessories, connect the inputs to the circuits under 

measurement. 

                             

  CAUTION. This probe is to carry out differential measurement 

                          between two points on the circuit under measurement. 

                                                    This probe is not for electrically insulating the circuit 

                                                    under measurement and the measuring instrument. 

                                                     

5. Appearance

The differential probe looks as follows. 

                         

       

                                                                           

  a. Output Cable                       

The BNC output connector and an auxiliary grounding terminal 

   

 are connected to the oscilloscope. 

b. Input Leads                         

The input leads of the differential probe connect to the 

                    

HV alligator clips that come with the probe. 

c. HV Alligator Clips      For safe test point connection 