4. Installation
a. Simply plug-in the BNC output connector to the vertical input of a general purposed
oscilloscope or other measurement instrument, and connects the auxiliary grounding
terminal to a proper ground. The measurement instrument must have a ground referenced.
b. Connect an appropriate power source to this probe and or enter the batteries, then turn it
on.
c. Select the proper attenuation ratio. When measuring signals below 700V switch the
attenuation ration to 1/100 in order to get higher resolution and less noise ratio. Otherwise,
set the attenuation ratio to 1/1000.
WARNING. To protect against electric shock, use only the
accessories supplied with this probe.
d. Using the appropriate probe accessories, connect the inputs to the circuits under
measurement.
CAUTION. This probe is to carry out differential measurement
between two points on the circuit under measurement.
This probe is not for electrically insulating the circuit
under measurement and the measuring instrument.
5. Appearance
The differential probe looks as follows.
a. Output Cable
The BNC output connector and an auxiliary grounding terminal
are connected to the oscilloscope.
b. Input Leads
The input leads of the differential probe connect to the
HV alligator clips that come with the probe.
c. HV Alligator Clips For safe test point connection