TV304 User’s Manual
18
Emissivity
:
To set the emissivity, press left or right key to adjust.
TCorrect
:
To setup system’s temperature correction.
(
normally set by manufacturer
)
Distance:
The distance between the machine and objects. Need to be set manually. (It is better to use the
actual distance within 10 meters, and farther than 0.5 meter)
Humidity:
Current environment relative humidity.
TReflect
:
To setup the system’s reflecting temperature.
(
normally set by manufacturer
)
8.2.3.3 Analysis
Figure 8.2.3.3 Analysis Setup
Temp alarm:
To turn on or off the temperature alarm. It is only efficient when doing area temperature
measurement. When it is turned on, and if the temperature is in the alarming temperature area,
the buzzer will automatically run and the region will be shown as the alarming color. The alarm
color can be set at ‘alarming color’ column.
Alarming Temp:
set an alarm temperature. This function is only efficient when temperature alarm is on.
Alarming Color:
there are 9 colors to choose. When choose ‘auto’ it will not change the original color of
the screen.
Isothermal color:
same as alarming color.
Isothermal temperature:
the temperature of the isothermal central point.
Isothermal width:
The value of a symmetry region around the central isothermal point. If the isothermal
temperature is 50°C, isothermal width is 2°C, then the isothermal range will be 49°C ~ 51°C.