![TeST TT AUT-MARSHALL-CBR Скачать руководство пользователя страница 16](http://html1.mh-extra.com/html/test/tt-aut-marshall-cbr/tt-aut-marshall-cbr_manual_1091833016.webp)
9
Connecting a DSU Readout device
9.1
Multiplex 50 machines fitted with transducers (instead of manual gauges) may be
connected to a transducer readout device called a
DSU
. This device allows real-time
display of transducer readings and storage of test data, which can then be
downloaded to a PC. The DSU also provides several other features such as
automatic overload shut-off, automatic logging start, automatic checking of speed
selection and peak force / displacement display.
There are 2 DSU connections on the rear
– one provides a serial link (9 way RS232
connector) and the other provides machine identification and remote shut-off
functions (round connector). The default serial port speed is 19200 baud. Please
see DSU manual.