GVT-427B User Manual

14

Table 1: Specifications (Cont.)

Frequency Response

(Reference : 1 kHz)

300uV range:

From 20 Hz to 200 kHz, 

 

±

3%

From 10 Hz to 500kHz, 

 

±

10%

Other ranges:

From 20 Hz to 200 kHz, 

 

±

3%

From 10 Hz to 1 MHz, 

 

±

10%

Distortion Factor

2 % of full scale at 1 kHz

Input Impedance Approximately  1 M

Input Capacitance

50 pF

Input Maximum

Voltage

(DC + AC peak)

300 V (at 300 

µ

V~ 1 V ranges)

500 V (at 3 V ~ 100 V ranges)

**DC isolation

Resistance

0.1 

Ω /100 

k

Ω 

±

10% (selectable)

Maximum Voltage of

**DC isolation

±

30 V (peak value)      (at 100 k

Ω 

isolation)

AC Output Voltage 0.1 V

rms

 

±

10% for each range of 1 kHz

(at full scale without load)

AC Frequency

Response

10 Hz to 1 MHz, 

 

±

3%

(reference: 1 kHz, without load)

Stability against Line

Voltage Fluctuation

Indication change with respect to line voltage

fluctuation of 

±

10% : is within 

±

0.5% of full

scale.

GVT-427B User Manual

15

AC Power

Requirement

115/230 Vac, 50/60 Hz

Table 1: Specifications (Cont.)

AC Power

Consumption

8.5VA or 5.0 Watts, Maximum

Fuse Replacement

for 115 Vac selected

T type, 0.3 A, 250Vac

Fuse Replacement

for 230 Vac selected

T type, 0.2 A, 250Vac

Operation

Environment

Indoor user,

Altitude up to 2000M

Installation Category             II

Pollution Degree

2

Operating Temperature :

+0

0

 C to +40

0

C, < 80% relative humidity

Dimensions

130 (W) x 210 (H) x 295 (D) mm

Weight

Approximately 2.84kg

*: Dual scales in different colors.

**: Between the input common and chassis.

Содержание GVT-427B

Страница 1: ...he product WARNING Warning statements identify condition or practices that could result in injury or loss of life CAUTION Caution statements identify conditions or practices that could result in damage to this product or other property The following symbols may appear in this manual or on the product DANGER High Voltage ATTENTION refer to Manual Protective Conductor Terminal Ground Earth Terminal ...

Страница 2: ...loured Blue or Black The wire which is coloured Brown must be connected to the terminal marked with the letter L or P or coloured Brown or Red If in doubt consult the instructions provided with the equipment or contact the supplier This cable appliance should be protected by a suitably rated and approved HBC mains fuse refer to the rating information on the equipment and or user instructions for d...

Страница 3: ...sis ground terminal is connected to the earth before inserting the power plug into the main supply 2 Maximum input voltage The voltmeter may be damaged if any input voltage exceeding the specified voltage is applied to it The specified voltage is determined by adding the peak value of the input signal and the superimposed DC voltage 300 V for the 300 µV ranges and 500 V for the 3 V to 100 V ranges...

Страница 4: ...ut signals when the meter is used as a preamplifier When the range selector switch is setting at 100 mV the output voltage will be approximately equal to the input voltage However when the range selector switch is setting to the next higher or lower voltage range the amplification factor is decreased or increased by 10 dB respectively The maximum voltage of DC isolation is 12 V peak value 10 Mode ...

Страница 5: ...GVT 427B User Manual 8 GVT 427B User Manual 9 ...

Страница 6: ...0 log P2 P1 If the impedance is at the place where P1 and P2 are equal to each other the ratio of power could be expressed as follows dB 20 log E2 E1 20 log I2 I1 GVT 427B User Manual 11 Decibel is originally the ratio of power as explained above However the logarithm of the ratio of other values ratio of voltage or current can also been called decibel For example If the input voltage of an amplif...

Страница 7: ...s section contains a table of GVT 427B characteristics Table 1 Specifications Model Specification GVT 427B Channels Two Indicating Meter 105mm in scale length 2 Pointers Black and Red Scale Values Vrms value of sinusoidal wave dB value with 1 V as 0 dBm dBm value with 1 mW Range Select Mode Independent tracking control mode for each channel range Voltage Measurement 12 ranges 300 µV 1 3 10 30 100 ...

Страница 8: ...ull scale without load AC Frequency Response 10 Hz to 1 MHz 3 reference 1 kHz without load Stability against Line Voltage Fluctuation Indication change with respect to line voltage fluctuation of 10 is within 0 5 of full scale GVT 427B User Manual 15 AC Power Requirement 115 230 Vac 50 60 Hz Table 1 Specifications Cont AC Power Consumption 8 5VA or 5 0 Watts Maximum Fuse Replacement for 115 Vac se...

Страница 9: ...er operational difficulties with your GVT 427B contact your Good Will representative for assistance WARNING To avoid electrical shock the power cord protective grounding conductor must be connected to ground WARNING To continued fire protection Replace fuse with the specified type and rating and disconnect the power cord before replacing fuse GVT 427B User Manual 17 Interchangeable Parts Interchan...

Отзывы: