background image

 [EN]  English - 

3

K345 Issue 4

To start - Power on or off

To turn the instrument on or off, press 

 (A1 - item [1]). The 

instrument does a self test and then shows the applicable 
data.
When the power is off, the last set of configuration options 
stays in memory. Refer to “Maintenance”.

To start - Set up the basic operation

Use the 

Set Up

 menu to set up the basic operation of the 

instrument. 

If there is additional data for a menu option, select 

Settings

 (

 

) to see the values that are set up. If 

necessary, adjust the values.

Table 1: 

Menu options - Set Up

To start - Select a task (Measure and/or simulate)

When the instrument is set up (Table 1), use the 

Select Task

 

menu to select the applicable task.

In Table 2/3, IDOS is a Universal Measurement Module 
(UMM). If you attach a UMM to the communications port 
(A1 - item [9]), the 

Select Task

 menu shows the applicable 

IDOS options.

Table 2: 

Menu options - Select Task

Table 3 shows all the one and two function operations that 
are available. If you attach a UMM, you can only use the 
options that include IDOS.

Table 3: 

Permitted 1 and 2 function operations

1

Menu:

Select Task

2

3

Menu:

Set Up

4

5

(Table 2)

(Table 1)

[

]/[  ]

Options

(If applicable)

Description

 

DPI 812 only. To add a series resistor into the mA 
circuit. You can then use this instrument together 
with a HART

®

 communicator to set up and 

calibrate HART

®

 devices.

... Scale To select the applicable international temperature 

scale: IPTS 68 or ITS 90.

 

To select and set up the backlight fa timer.

Additional data: Select Settings 

(

 

)

 

To select and set up the power off fa timer. 

Additional data: Select Settings 

(

 

)

 

To show the battery level (%).

 

To set the display contrast (%).

▲ 

Increases %, 

▼ 

decreases %

 

To set the time + date. The calibration facility uses 
the date to give service and calibration messages. 

 

To set the language option.

 

To calibrate the instrument.

Additional data: Refer to “Calibration”.

 

To select and show the applicable status data. 
(Software Build, Calibration Due date, Serial 
Number, IDOS Information).

1

Menu:

Select Task

2

3

Display:

RTD output

(Table 2/3)

Sk1 = Edit
Sk2 = Settings

Options

(If applicable)

Description

 RTD 

or

 

Ohms

An input measurement task: 

 

RTD - Measure RTD temperature  OR

 

Ohms - Measure RTD resistance.

RTD or 

 

Ohms

An output task: 

 

RTD - Simulate RTD temperature  OR

 

Ohms - Simulate RTD resistance.

mA

DPI 812 only. A mA measurement task. 

mA(24V)

DPI 812 only. A mA measurement task + the loop 
power supply is on. 
DPI 812 only. A switch test. 

 

IDOS

UMM only. An IDOS measurement task. 

 UPM

 

only. A leak test. 

To set up the way the instrument works.

Additional data: Refer to: Set Up 

(Table 1).

Function

RTD

RTD

Ohms

Ohms

 

IDOS

(1)

(1)

(1)

(1)

(1)

mA

(1) x

(2)

x

(2)

(2)

mA(24V)

(1) x

(2)

x

(2)

(2)

x

x

(2)

x

(2)

(2)

x

x

x

x

x

(2)

IDOS

(1) (2)

(2)

(2)

(2)

x

= DPI 812 only

Содержание Druck DPI 811

Страница 1: ...Druck DPI 811 812 RTD calibrator and RTD loop calibrator User manual K345 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot com...

Страница 2: ...K345 Issue 4 A1 A2 DPI 812 A3 10 1 3 4 2 5 6 7 9 8 A2 A3 11 B1 12 10 14 15 13 20 21 22 25 17 18 26 19 27 16 23 24 B1 Customer service Visit our web site www gesensing com...

Страница 3: ...s uses Intelligent Digital Output Sensor IDOS technology to give instant plug and play functionality with a range of Universal Measurement Modules UMM Example the Universal Pressure Module UPM The DPI...

Страница 4: ...ion above it on the display Item 25 Example Settings 8 Display Refer to A3 9 SENSOR PC Communications port Use to connect a Universal Measurement Module UMM or a RS232 cable 10 RTD connectors To measu...

Страница 5: ...se this instrument together with a HART communicator to set up and calibrate HART devices Scale To select the applicable international temperature scale IPTS 68 or ITS 90 To select and set up the back...

Страница 6: ...select and set up a tare value for the measurement task a specified value or the reading on the display The instrument subtracts a positive tare value and adds a negative tare value Additional data S...

Страница 7: ...ional Additional data Table 6 Select Settings Step To select and set up values for the Defined Step output Example output cycle Auto Repeat Optional Additional data Table 6 Select Settings Options If...

Страница 8: ...adjust the Set Up Table 1 2 Select the RTD or Ohms input task from Select Task Table 2 3 and if necessary adjust the Settings Table 4 Figure 1 Example configuration To measure the temperature or resis...

Страница 9: ...ransmitter calibration DPI 812 only To calibrate a transmitter 1 Connect the instrument Figure 3 4 and if necessary adjust the Set Up Table 1 2 Select the applicable calibration task from Select Task...

Страница 10: ...the top right hand corner 3 Supply the output values to the system Table 7 Example Nudge output a Use Edit to set a value less than the switch value b Use the buttons to change the value in small incr...

Страница 11: ...Settings Table 4 5 2 If necessary do a zero correction Table 4 To measure pressure with another operation Figure 8 use the same procedure Figure 8 Example configuration To measure pressure and temper...

Страница 12: ...set the new calibration date for the instrument Calibration Procedures Ohms input 1 Let the equipment get to a stable temperature minimum 5 minutes since the last power on 2 Use the calibration menu...

Страница 13: ...nt get to a stable temperature minimum 5 minutes since the last power on 3 Use the calibration menu Table 9 to do a three point calibration FS Zero and FS The display shows the applicable instructions...

Страница 14: ...850 0 90 0 50 Pt100 385 IEC 751 328 1562 200 850 0 45 0 25 Pt200 385 IEC 751 328 1562 200 850 1 08 0 60 Pt500 385 IEC 751 328 1562 200 850 0 72 0 40 Pt1000 385 IEC 751 328 752 200 400 0 36 0 20 D 100...

Отзывы: