background image

7

Dimensions L x W x H 

360 x 180 x 150 mm (14.2 x 7.1 x 5.9”)

Weight 

12 kg (26.5 lbs) approx.

Cable length to NSG 

2 m (79”)

Input cable 

NSG 3000 standard cable to be used

 

Part designation 

Function

EUT power in switch 

Switches on/off the EUT power supply

16 AT fuse 

Protects the EUT power supply line

EUT power in plug 

This is where the EUT power needs to be 

 

 

applied – use NSG 3000 supplied cable 

 

 

and conect it to power source (mains)

EUT power out cable 

Connect this cable to the EUT supply 

 

 

input port of NSG 3000

Red selector 

To switch variable voltage to 0 or 40% or 

 

 

70 or 80% of Uin

 

1.2.1 Parts description

1.3 Installation

 

The equipment should be switched off during installa-
tion and interconnection.

  Connect INA 6501 - EUT power out to NSG - EUT power input.
  Connect  INA  6501  -  EUT  power  in  to  mains  using  EUT  power  in  cable 

 

  (delivered with NSG).

Содержание INA 6501

Страница 1: ...1 MANUAL AUTOMATED STEP TRANSFORMER INA 6501 INA 6502 USER MANUAL 601 262C...

Страница 2: ...Manual automated step transformer INA 6501 INA 6502 MANUAL AUTOMATED STEP TRANSFORMER INA 6501 INA 6502 601 262C USER MANUAL...

Страница 3: ...1 2 Technical specifications INA 6501 6 1 2 1 Parts description 7 1 3 Installation 7 2 Automated step transformer INA 6502 9 2 1 Circuit diagram INA 6502 10 2 2 Technical specifications INA 6502 10 2...

Страница 4: ......

Страница 5: ...in case of use in standing posi tion as the stability is limited The cabling connecting INA 6501 to mains and NSG presents a risk of being unvoluntarly caught by the users which could cause the INA 6...

Страница 6: ...0 A less than 5 Output current capability at 16 Arms at 100 output 230 V input voltage 20 Arms at 80 output for 5 s 23 Arms at 70 output for 3 s 40 Arms at 40 output for 3 s Voltage selection Front pa...

Страница 7: ...is where the EUT power needs to be applied use NSG 3000 supplied cable and conect it to power source mains EUT power out cable Connect this cable to the EUT supply input port of NSG 3000 Red selector...

Страница 8: ...00 generator Select the required variable voltage using the rotary switch on the INA 6501 Because of the capacitors in the internal coupler of the CDN earth leakage currents of up to 4 A can occur in...

Страница 9: ...A 6502 is fitted with carrying handles as part of its overall excellent ergonomic design which allows easy handling Further the unit may be used in any of three operating positions laying or standing...

Страница 10: ...output for 5 s 23 Arms at 70 output for 3 s 40 Arms at 40 output for 3 s Voltage selection User interface or WIN 3000 software 2 1 Circuit diagram INA 6502 L 1 1 3 2 2 EUT Power IN N C PE N EUT Power...

Страница 11: ...m 79 Input cable NSG 3000 standard cable to be used Control cable 2 m 25 way sub D twisted pair shielded included in delivery Part designation Function Mains supply in socket For instrument supply inc...

Страница 12: ...out to be terminated by interlock plug or linked to another accessory to X1 plug Plug X1 NSG 3000 system interface in to be connected to master controller or to another accessory to X2 plug Power LED...

Страница 13: ...mains using EUT power in cable delivered with NSG or CDN Because of the capacitors in the internal coupler of CDN earth leakage currents of up to 4 A can occur in the EUT power supply network The tes...

Страница 14: ...nterprise Shanghai Co Ltd Beijing Branch T 86 10 8526 2111 F 86 10 8526 2141 chinasales teseq com France AMETEK SAS T 33 1 30 68 89 00 F 33 1 30 68 89 99 info france ametek com Germany Teseq GmbH T 49...

Отзывы: