38
RS-232C Commands
5.5.2.2 Handler & Probe Configuration Commands
SOLENOID MODE?
HANDLER MODE SENSOR
HANDLER MODE TIMER
HANDLER MODE OFF
PROBE MODE SENSOR
PROBE MODE TIMER
PROBE MODE OFF
SOLENOID MODE?
A query command to see the Handler & Probe operation modes.
Type
Query
Input
SOLENOID MODE?
Response
Handler:<mode>, Probe:<mode>
●
<mode>: {OFF | TIMER | SENSOR }
Example
// Below is how to change the Probe Mode from Off to Timer.
SOLENOID MODE?
Handler:OFF, Probe:OFF
PROBE MODE TIMER
OK
It is NOT recommended operating fixture with the Timer Mode.
Without sensors, the lid may close even when the fixture operation is incomplete, thus
causing damage to the product or DUT.
HANDLER MODE SENSOR
A command to set the Handler in Sensor Mode.
Type
Command
Input
HANDLER MODE SENSOR
Response
OK