KATC-C1 Instructions
Page 6
04/14
Operation
Description of Controls
ON/OFF Switch
Controls 120 VAC power to the KATC-C1.
REFERENCE TRANSFORMER Switch
Used to configure the KATC-C1 for use with reference (precision) transformers with 1 or 5 ampere
secondaries.
Do not change this switch with current applied to the KATC-C1.
TRANSFORMER-UNDER-TEST Switch
Used to configure the KATC-C1 for testing transformers with 1 or 5 ampere secondaries.
Do not
change this switch with current applied to the KATC-C1.
RATIO ERROR Switch
Allows the ratio error of the transformer-under-test to be displayed in Percent Error (%E) or Ratio
Correction Factor (RCF). [%E = 100 * (RCF-1)]
PHASE ERROR Switch
Allows the phase angle error of the transformer-under-test to be displayed in Minutes (MIN) or Milli-
radians (MR). [1 Minute = 0.29089 Milliradians]
50/60 Hz LEDs
Indicate whether the KATC-C1 has detected the applied test current to be 50 or 60 Hz. If no test
current is applied, the LEDs default to 60 Hz.
MEASURE Push-button
Toggles the KATC-C1 between MEASURE and RUNUP modes.
MEASURE mode should be selected when the desired test current has been reached. Selecting
MEASURE mode causes the KATC-C1 to begin taking measurements using the optimum meas-
urement range. Measurements can be taken at any current greater than approximately 4% of the
selected transformer-under-test secondary current. Selecting MEASURE mode erases any data
previously saved by pressing the HOLD/RUNDOWN push-button.