INSTALLATION
9
+1 215 785 2338
2.3.2 Airflow
Take note of the Instrument’s airflow
as seen in the illustration. This always
for a table top, never on a rack. Please
allow enough airspace at the side with
at least half a meter for an effective
airflow.
CAUTION
Damage caused by overheating may occur if the area around the air intake is restricted, the
intake air is too warm, is interfered, or the air filter becomes clogged.
The inlet and exhaust holes must be clear of obstruction.
The air entering the instrument must be between 5 °C
and 35 °C. Make sure that exhaust from another
instrument is not directed into the fan inlet.
Check and clean the air filter every 30 days or more
frequently if the Instrument is operated in a dusty
environment. Refer to §5.3 for additional info.
2.4 Main Power Supply
The Instrument can be powered by plugging it to a 120V-Single Phase AC line. An AC line power
cord is provided.
1-
Power Switch
2-
5A Fuse
3-
3-prong Single Phase
120V AC Line