Storage temperature : 

 

and humidity           –10

 to 60

 (14

 to 140

), 

                                            below 70%RH. 

Dimensions :   

170(L)×52(W)×38mm(H) 

6.7(L)×2.1(W)×1.5(H) inches. 

Weight :   

Approx. 180g with battery. 

Accessories :   

instruction manual, carry case, 
battery. 

 

3-2 Electrical Specifications 

‰

 Infrared   

Measuring range :    -20

 to 500

 (-4

 to 932

Resolution :   

1

 , 1

 

Accuracy :   

± 3% reading or ± 3

 or 6

Temperature  

coefficient : 

  0.1 times the applicable accuracy   

                                    specification per 

 from 0

 to 18

 

                  and 28

to 40

(32

to 64

and 82

 

                                    to 104

). 

Responding time :    once per second. 

Spectral Response :   

6 ~ 14um. 

Field of view :   

8:1 ; optics ratio with a 1” min target. 

Emissivity :   

  0.1 ~ 1.0 

Sighting :   

Laser marker 1mw (class 2). 

Sensor :   

Thermopile. 

‰

 Type K   

Measuring range :   -50

 to 1333

 (-58

 to 1999

). 

Resolution :   

0.1

 , 1

 , 0.1

 , 1

Responding time :   once per second. 

Basic accuracy :    (@ 23±5

  calibration) accuracy are ± 

(…% of r degree) at 18

 to 

28

  with relative humidity up 80%. 

 

Function Resolution 

Range 

Accuracy 

-50

℃〜

0

 

±

( 0.2%rdg + 1.0

0.1

 

0

℃〜

200

 

±

( 0.1%rdg + 0.8

 

1

 200

℃〜

1333

 

±

( 0.2%rdg + 2

-58

℉〜

32

 

±

( 0.2%rdg + 2

0.1

 

32

℉〜

200

 

±

( 0.1%rdg + 1.6

 

1

 200

℉〜

1999

 

±

( 0.2%rdg + 3

Temperature Coefficient: 

0.1 times the applicable accuracy specification per 

 

from o

 to 18

 and 28

 to 40

 ( 32

 to 64

 and 

82

 to 104

 ). 

 

Note :   

The basic accuracy specification does not include the 
error of the probe.    Please refer to the probe 
accuracy specification for additional details. 

 

 

Содержание TES-1322

Страница 1: ...r It can be used to measure the temperature of objects surface that is improper to be measured by traditional contact thermometer such as moving object the surface with electricity current or the objects which are uneasy to be touched Ⅰ SAFETY INFORMATION 1 Read the following safety information carefully before attempting to operate or service the meter 2 use the meter only as specified in this ma...

Страница 2: ...0 max c Operating Ambient 0 40 5 Maintenance Clearing a Repairs or serving aren t covered in this manual should only be performed by qualified personnel b Periodically wipe the case with a dry cloth Don t use abrasives or solvents on this instrument c When serving use only specified replacement parts 6 Safety symbols Comply with EMC 7 U S Pat No Des 448 314 2 Ⅱ FEATURES Selectable Back light LCD d...

Страница 3: ...o with a 1 min target Emissivity 0 1 1 0 Sighting Laser marker 1mw class 2 Sensor Thermopile 4 Type K Measuring range 50 to 1333 58 to 1999 Resolution 0 1 1 0 1 1 Responding time once per second Basic accuracy 23 5 calibration accuracy are of reading degree at 18 to 28 with relative humidity up 80 Function Resolution Range Accuracy 50 0 0 2 rdg 1 0 0 1 0 200 0 1 rdg 0 8 1 200 1333 0 2 rdg 2 58 32 ...

Страница 4: ... mode and mode settings selection a Press MODE button to select measuring mode Infrared thermometer Type K thermometer b Press MODE button for 3 seconds entering into the mode settings 6 Type K input connector 7 Laser aperture 8 Focal lens Ⅴ LCD DISPLAY DESCRIPTION 1 Low battery mark 6 Laser indicator 2 Maximum 7 Hold function 3 Negative polarity 8 Hi Alarm 4 Measure value 9 Lo Alarm 5 Unit 10 SET...

Страница 5: ...lues Metals cont SURFACE EMISSIVITY Zinc oxidized 0 1 Galvanized iron 0 3 Tin plated steel 0 1 Gold polished 0 1 Silver polished 0 1 Chromium polished 0 1 Emissivity Values Non Metals Refractory Building Materials Red brick rough 0 75 to 0 9 8 Fire clay 0 75 Asbestos 0 95 Concrete 0 7 Marble 0 9 Carborundum 0 85 Plaster 0 9 Alumina fine grain 0 25 Alumina coarse grain 0 45 Silica fine grain 0 4 Si...

Страница 6: ...95 10 Ⅶ TEMPERATURE MEASUREMENT 1 Infrared measuring a Infrared measuring Press MEAS button to power on the meter and start measuring Release MEAS button to stop measuring and auto hold the reading the meter will be off automatically after 15 secs b Under the Infrared measuring mode press hold MEAS button and press MODE button to select MAX measurement General measuring mode 2 Continuous infrared ...

Страница 7: ...gs of infrared measurement if K doesn t appear in LCD 12 1 Parameter settings for Infrared measuring Under the infrared parameter setting mode LCD will show press MODE button to select in turn for setting ε ALMÇ ÈALM press MODE button again to escape the setting mode and return to the general measurement Note In the setting mode if users don t push any button in 15 seconds then it will escape from...

Страница 8: ...ser light function will be disabled Ⅸ BATTERY REPLACEMENT 1 As battery power is not sufficient LCD will display replacement with one new battery type 9V is required 2 Open battery cover then take out the battery from instrument and replace with a new 9 Volt battery and place the battery cover back 14 TES ELECTRICAL ELECTRONIC CORP 7F No 31 Lane 513 Rui Guang Road Neihu Dist Taipei Taiwan R O C Tel...

Отзывы: