1.Place the Bionova
®
MiniPro Auto-reader on a firm surface,
free from vibrations, away from direct sunlight, currents of hot
or cold air, chemical and corrosive or flammable substances.
Do not place the Auto-reader in a way that disconnection of the
power supply’s AC plug from mains could be difficult. Leave
a space of at least 10 cm from the Auto-reader to the closest
wall. Do not move the Auto-reader periodically or during its use.
Connect the Auto-reader to a secure and stable electrical mains
outlet socket.
Bionova
®
MiniPro Auto-reader do not represent photobiological
risk and do not generate dangerous optical radiation in any of
its normal operation conditions as per the requirements of IEC
62471 Standard.
Designed under Quality Management System standards ISO
13485:2016/NS-EN ISO 13485:2016.
The following incubation programs are available for your
Bionova
®
MiniPro. Verify which incubation programs are available
for your country checking the
Indications for Use
section.
Product Compatibility and
Incubation programs
Bionova
®
MiniPro Incubation Programs
Temperature
selection
60 ºC
Time selection
PRO
(10 min.)
Incubation Program
PRO at 60°C
Select the appropriate incubation program before
the incubation of any Indicator.
We develop new Indicators for extending the features of our
Auto-readers regularly. Please visit
www.terragene.com/
compatible
or scan the QR code below to check the latest
compatible Indicators developed for your Auto-reader.
To select the appropriate incubation program for the Indicator you
are planning to use, follow each Indicator’s Instructions for use.
Bionova
®
MiniPro Auto-reader compatible indicators
Serial/Batch Number
For easy identification, each Bionova
®
MiniPro Auto-reader has
a distinctive serial and batch numbers. The number can be
found printed on two labels, one placed at the bottom of the
Auto-reader, and one on its packaging. Please, register your
serial and batch numbers in this manual for future reference:
Serial number:
Batch number:
Instructions for use
Do not wet or heat the device. If liquid is spilled on the
Auto-reader, disconnect it and follow the instructions
on the
Cleaning and maintenance
section.
2.Power on your Bionova
®
MiniPro by connecting the power
supply’s AC plug to mains and then connect the plug at the
other end of power supply to the rear of the Auto-reader.
Before power on, verify that all incubation positions
are empty.
If the Blue Attention light starts blinking after
powering on the Auto-reader, check that the
printer’s door is tightly closed and that the printer
has paper, if not follow the instructions on the
Replacement of the paper roll
section.
3.Set the Auto-reader internal clock to correspond to your local
time. See configuration methods in the Time setting mode
section.