Terra Universal Ductless Exhaust Fume Hood Скачать руководство пользователя страница 3

Installation and Operating Guide  

 

High- Ductless Exhaust Fume Hood 

© 

Copyright 2016 

Terra Universal Inc.  All rights reserved • Revised December 2016 • Document No. 1800-72 

Terra Universal, Inc.  •  TerraUniversal.com  •  800 S. Raymond Ave.  •  Fullerton, CA  92831  •  TEL: (714) 578-6000  •  FAX: (714) 578-6020 

 

 

 

1.0

 

Description 

 

Terra's Ductless Exhaust Fume Hoods provide safe removal of a broad range of organic chemical fumes without the 

need for cumbersome ductwork. 

 

These high-capacity hoods feature dual 3000-RPM impellers (each rated for 740 CFM @ 0" static pressure) to ensure 

a safe exhaust flow speed of 100 fpm under filter load. Bonded activated charcoal filters eliminate voids in the filter 

medium and minimize particle sloughing. For cleanroom operation, an exit HEPA filter, rated 99.99% efficient @ 0.3- 

micron particle size, may be specified. 

 

Terra's standard model (No. 2900-20) is ideal for most benchtop applications. The all-polypropylene construction is 

resistant to most chemicals. This model features a built-in air Vaneometer™ Air Velocity Meter for easy monitoring of 

the exhaust air speed. 

 

Terra's High-Clearance Chemical Fume Hood (Nos. 2900-30 through 2900-33) provides over 28" of vertical clearance 

with the access shield open, making it ideal for distillation and other operations with large glassware or equipment. It 

is available in several hood materials and includes a spill tray made of white polypropylene. 

 

Vapor-Sealed, Fluorescent Lighting

 

 

In order to avoid potential fire or explosion hazards, the Ductless Exhaust Fume Hood includes vapor-sealed lighting 

as a standard light fixture. 

 

 

2.0

 

Installation 

 

Carefully uncrate the Ductless Exhaust Fume Hood and visually inspect for damage. Any damage should be reported 

according to the procedures outlined in the shipping agreement. 

 

The hood is normally shipped fully assembled, with filter(s) installed inside the filter housing and both filter and blower 

housings securely clamped in place. 

 

To  operate  the  unit,  simply  place  the  hood  on  a  sturdy,  level  work  surface,  plug  the  blower  power  cord  into  a 

grounded 115/230VAC, 50/60Hz power source, and turn it on (the blower must be ordered for the proper voltage 

supply). 

 

 

 

Содержание Ductless Exhaust Fume Hood

Страница 1: ... Document No 1800 72 Ductless Exhaust Fume Hood Copyright 2016 Terra Universal Inc All rights reserved Revised December 2016 Terra Universal Inc TerraUniversal com 800 S Raymond Ave Fullerton CA 92831 TEL 714 578 6000 FAX 714 578 6020 ...

Страница 2: ... and filter integrity must be monitored and tested frequently Terra Universal assumes no liability for damage or injury that may result as a consequence of improper filter selection use or testing NOTE Terra Universal also offers toxic gas detection systems for testing the activated carbon filter for saturation Please contact a Terra representative for more information or follow the link below htt...

Страница 3: ...tures a built in air Vaneometer Air Velocity Meter for easy monitoring of the exhaust air speed Terra s High Clearance Chemical Fume Hood Nos 2900 30 through 2900 33 provides over 28 of vertical clearance with the access shield open making it ideal for distillation and other operations with large glassware or equipment It is available in several hood materials and includes a spill tray made of whi...

Страница 4: ...y two years depending on use As these filters clog you will notice a reduction in air speed and noisier blower operation To replace filters from inside the work area The filter s are held securely in place by two plated steel pawls along the front edge of the filter housing Rotate each black knob 90 to release the filter s and then carefully lower them out of the housing Insert the replacement fil...

Страница 5: ... x 0 5 Weight 104 lbs Electrical 120 240VAC 50 60Hz 20A 1Ph Individual Blower Specifications Table Horsepower HP Power Use 196 W Max Amps 1 96 A Max Airflow 598 CFM Fan Speed 3104 RPM Fan Wheel Type Backward centrifugal Impeller High impact polypropylene external rotor Insulation Class B Enclosure IP44 Housing Heavy gauge galvanized sheet metal Motor Bearings Permanently sealed self lubricating ba...

Страница 6: ...or separately For Terra s Policy and Procedures for Returning Goods see Terra s Internet site www TerraUniversal com Generally customers can improve the chance of collecting on a freight claim by following these procedures 1 formally requesting that the carrier inspect the shipment immediately upon suspecting damage or shortage to verify condition 2 notifying the carrier upon discovery of conceale...