
Non-Contact Forehead IR Thermometer is designed for body surface and forehead temperature
measurement for infants and adults without contact to human body.
Non-Contact Forehead IR Thermometer can also be used to measure the temperature of a baby-
bottle or bath, or room temperature (by using the Surface Temp function).
Pyrometer measuring range:
- ambient temperature: 0,0...60,0°C
- body temperature: 32,0...42,5°C
Display Resolution:
0,1°C
Accuracy:
±0,3°C
Measuring distance:
5...15cm
Display:
LCD with backlight
Automatic Data Hold & Auto power off:
after 7 sec of idleness
Unit selection:
°C / °F
Buzzer:
indicating about a measurement and alarm threshol exceeding
Functions:
- last 32 measurements memorizing
- measurements calibration in: -3...3°C range
Dimensions/weight:
128x74x36 mm / 125g
Power supply:
DC 3V ( 2 x 1,5V battery )
Operation environment:
0°C÷50°C, <85% RH
Equipment:
batteries, case, user manual
This appliance conforms to the following standards:
EN 980: Graphic symbols for use in the labeling of medical devices
EN 1041: Information supplied by the manufacturer with medical devices
EN 60601-1: General requirements for safety
EN60601-1-2: General requirements for safety Collateral standard Electromagnetic
compatibility
3. USE
Aim towards the forehead from a distance of 5 cm (2in), press the measuring key, the
temperature is displayed immediately. Making sure there is no hair, perspiration, cosmetic or
cap covered on the forehead.
When the room temperature is significant difference, or there is perspiration on the forehead,
You can take the temperature behind the ear lobe. Making sure there is no hair, perspiration,
cosmetic or cap covered.
Attention !!!
- For the first use or when inserting new batteries wait between 10 minutes for the warm-up of the
apparatus and when inserting the new batteries.
- If the device is not used for a long time, once you turn it on again, the device will test the room
temperature first and will delay turning on for one or two seconds,
1. GENERAL DESCRIPTION
2. TECHNICAL SPECIFICATIONS
4.BODY TEMPERATURE MEASUREMENT (BODY)
Body temperature measuring range is 32,0
The temperature of the human body varies throughout the day. It can also be influenced by
numerous external factors: age, sex, type and thickness of skin...
Temperature according to the age:
0-2 years:
36,4
...38,0
Oral:
35,5
...37,5
°C - 42,5°C. If the measurement value is not within a
range, the display can indicates following communicates:
HI
- when the temperature exceeds 42,5°C
Lo
- when the temperature is below 32,0°C
°C
°C
3-10 years:
36,1°C...37,8°C
11-65 years:
35,9°C...37,6°C
> 65 years:
35,8°C...37,5°C
Normal temperature in different parts of body:
Forehead:
35,5°C...37,6°C
Ear:
35,8°C...38,0°C
°C
°C
Axilliary::
34,7°C...37,3°C
5. DISPLAY
1. Ambient temperature measurement
2. Body temperature measurement
3. Digital readout
4. Battery Condition
5. Measurement number
6. Saved measurement
7. Unit
8. Alarm Symbol
6. DATA READOUT
Turn on the unit with button no.7 “ measurement trigger”. Then pass between the stored data
pressing 3 or 4 button.